Thermo Scientific Nicolet iS5 LS User Manual page 44

Table of Contents

Advertisement

3
Maintaining the Instrument
40
Nicolet iS5 LS User Guide
14. Install the desiccant module.
Insert the desiccant module into the instrument so that the text on the humidity indicator
can be read by someone standing in front of the instrument. Make sure the module is
seated properly over the O-ring and then tighten the two captive screws. (If the captive
screws are worn or damaged, replace them with the same type of screws from the
desiccant replacement kit.)
Figure 33. Nicolet iS5 LS analyzer with desiccant module installed
Desiccant module
Install
desiccant module
Thermo Scientific

Advertisement

Table of Contents
loading

Table of Contents