Calibrating the Laser Frequency
Thermo Scientific
Laser calibration verifies the known frequency of the instrument laser. The laser frequency
may drift slightly over time, particularly in locations that have large temperature swings or
excessive vibration. After you power up the instrument, allow approximately 6 hours for the
instrument to reach thermal equilibrium before you calibrate the laser frequency. Repeat laser
calibration every three months.
NOTICE Always
align the instrument optics
calibrate the laser frequency.
Calibrate the laser frequency immediately:
• after you have changed the source
• if the instrument failed the Wavelength Accuracy portion of the
• Before you align the instrument or calibrate the sample stage
Before you calibrate the laser frequency, make sure the sample windows are clean and dry.
The sampling arm should be down.
To continue from one screen to the next in the workflow, click the
To calibrate the laser frequency
1. From the qPro main window, click System Tools and then click Laser Calibration.
The workflow shows its historical results.
2. Make sure the sampling arm is closed and click continue.
3. Make sure Start a new laser calibration is selected and then click continue.
Follow the instructions in the workflow. When the workflow is completed, a pass or fail
result is displayed. If laser calibration fails, consult the
topic in the Troubleshooting section.
4. Click continue to return to System Tools.
5. Click the Home button at the top of the qPro window to return to the home screen.
After laser calibration is completed,
Related Topics
Aligning the Instrument Optics
Calibrating the Sample Stage
Maintaining the Instrument
Troubleshooting the Instrument
and
calibrate the sample stage
Laser frequency calibration failed
align the instrument optics
Maintaining the Instrument
after you
Calibration Check
button.
and
calibrate the sample
Nicolet iS5 LS User Guide
3
stage.
27

Advertisement

Table of Contents
loading

Table of Contents