Calibration Check
Thermo Scientific
5. If your instrument passed the performance test, click continue to complete the workflow.
If your report includes a red "X," rerun the Daily Performance Test workflow. If the
instrument fails again, the resolution depends on which test has the red X. For more
information, refer to the
the instrument user guide.
6. To print the report, click the Print button.
Note To retrieve data from a Daily Performance Test workflow, use Data Hub.
Related Topics
Calibration Check
Instrument Performance Tests
The Calibration Check workflow uses hexadecane to verify wavelength accuracy and sample
pathlength. We recommend that you run the Calibration Check workflow at least once a
week. The test can be run more often as needed to satisfy your company's internal protocols.
Before you run the calibration check, the instrument should be
and the qPro Operator software should be running. The sampling arm should be closed.
To continue from one screen to the next in the workflow, click the
Materials required:
• Small volume of 99% hexadecane (available from us)
• Small, clean dropper bottle or pipette
• Clean, dry, lint-free laboratory wipes (unscented and lotion-free)
• Powder-free nitrile gloves
CAUTION Avoid chemical exposure. Hexadecane is a chemical irritant. When handling
hexadecane:
• avoid contact with skin and eyes
• avoid inhalation of vapor or mist
• wear gloves that provide adequate protection
Performance test failed
topic in the Troubleshooting section of
Operating the Instrument
powered on
for 4 to 6 hours
button.
Nicolet iS5 LS User Guide
2
15