Collecting Spectra
Thermo Scientific
Figure 7.
Nicolet iS50 Advanced Diagnostics window
7. Choose Run Selected Test.
The software collects background and sample scans to generate a 100% transmission
spectrum and calculates the spectrum's signal to noise ratio. The numerical result is added
to the Actual column in the diagnostics window and a Pass or Fail result appears in the
Status column. If the Actual noise value is less than the Expected (or maximum allowable)
value, the test will report a pass result.
8. Close the Bench Diagnostics window.
Note If the AEM fails the performance test, consult the
contact our local service representative for assistance.