Collecting Spectra

Thermo Scientific
Figure 7.
Nicolet iS50 Advanced Diagnostics window
7. Choose Run Selected Test.
The software collects background and sample scans to generate a 100% transmission
spectrum and calculates the spectrum's signal to noise ratio. The numerical result is added
to the Actual column in the diagnostics window and a Pass or Fail result appears in the
Status column. If the Actual noise value is less than the Expected (or maximum allowable)
value, the test will report a pass result.
8. Close the Bench Diagnostics window.
Note If the AEM fails the performance test, consult the
contact our local service representative for assistance.
Related Topics
Troubleshooting
This section explains how to configure your AEM workflow and collect spectra. These topics
are covered:
Preparing the AEM for Sampling
Configuring Your AEM Workflow
Measuring a Sample with the AEM