Injectors; Lighting Specifications; Room Light Distribution - GE Innova IGS 6 Preinstallation Manual

Table of Contents

Advertisement

NOTICE
To power the AW splitter, use the appropiate cable out of lot P/N 2385173.

5.11 Injectors

The injector is provided with an additional ground cable (P/N #2135737) to meet equipotentiality
requirement at patient vicinity.
Injector emergency power off:
• if no emergency power off is required for the injector, the injector can be powered from a wall outlet
socket.
• if emergency power off circuit must include the injector, ensure the power for the injector is via the
PDB.
Remote Injector (rack mount)
A power cord is supplied with the injector.
110 VAC or 230 VAC are supplied from the main disconnect room device. The injector must be connected
to this supply.
Pedestal Injector
A power cord is supplied with the injector.
In all cases (110 VAC or 230 VAC), it will be connected to a wall outlet near the operator location, fed by
the main disconnect room device.

5.12 Lighting Specifications

5.12.1 Room Light Distribution

Requirements for lighting
Requirement for lighting concern the following, general, light–technique characteristics:
• Illuminator level.
• Lighting distribution.
• Preventing the operator from being dazzled by the light (by direct light sources or by reflection on
bright objects).
The Illumination level must be compliant with established lighting technical rules and be as constant as
possible.
Technical room, operating room and control room shall be provided with appropriate lighting in the
maintenance area (maintenance area to be considered are service workplaces). It corresponds to service
areas as defined for any of the product components.