Electrical System; General - Nikon Eclipse Ni-E Instruction Manual

Upright motorized microscope
Hide thumbs Also See for Eclipse Ni-E:
Table of Contents

Advertisement

2

Electrical System

General

2.1
■ Power
Problem
There is no power even
though the power switch
is on.
A motorized accessory
does not operate.
Contact arm LED is
blinking.
■ Illumination
Lamp does not light.
Lamp does not turn off.
Chapter 2
Cause
The power cord is not connected, or is
connected improperly. Or, the cable
between the control box and the main body
is not connected, or is connected
improperly.
A contact arm is not used.
A connector box is not used.
The motorized unit connected to the
contact arm is attached incorrectly.
The motorized unit connected to contact
arm is not in correct combination.
The motorized unit connected to contact
arm is not in correct combination.
There is no power supplied.
The lamp has burned out.
The lamp is not attached.
Operation is disabled.
Troubleshooting
Connect the cable properly.
(→See Chapter 1 "3 Assembly Method - 25 Connect
the motorized unit cable" in the
"Assembly/Maintenance".)
Use a contact arm.
(→See Chapter 1 "3 Assembly Method - 8 Attach
the contact arm/standard arm" in the
"Assembly/Maintenance".)
Use the connector box.
(→See Chapter 1 "3 Assembly Method - 9 Attach
the connector box" in the "Assembly/Maintenance".)
Attach the motorized unit correctly.
(→See Chapter 1 "3 Assembly Method" in the
"Assembly/Maintenance".)
Attach the motorized unit in the correct
combination.
(→See Chapter 1 "3 Assembly Method -
Introduction: Motorized Device Combination" in the
"Assembly/Maintenance".)
Attach the motorized unit in the correct
combination.
(→See Chapter 1 "3 Assembly Method -
Introduction: Motorized Device Combination" in the
"Assembly/Maintenance".)
Plug in the power cord.
(→See Chapter 1 "3 Assembly Method - 26 Connect
the power cord" in the "Assembly/Maintenance".)
Replace the lamp with the specified type.
(→See Chapter 3 "1 Replacing the Lamp" in the
"Assembly/Maintenance".)
Attach a designated lamp.
(→See Chapter 1 "3 Assembly Method - 6 Attach
the lamp" in the "Assembly/Maintenance".)
Change the setting.
(→See Chapter 3 "23 Operation on DS-L3 - 23.1
Setting Up the Microscope - (3) Configuring the
Button Functions - (3-1) Ni-E microscope button" in
the "Operation".)
45