Chart-out signal conditions:
Signal
Flow Rate
System Pressure
Composition
Sample manager I/O signal connectors
The rear panel of the sample manager includes a removable connector that
holds the screw terminals for I/O signal cables. This connector is keyed so that
it can receive a signal cable inserted only one way.
A contact closure output connection (Inject Start Out) from the
Requirement:
sample manager is required to trigger a mass spectrometer.
Sample manager I/O signal connectors:
1 2 3 4 5 6
For electrical specifications, see
Parameter Setting at 0
Volts (Minimum)
0.000 mL/min
–345 kPa (–3.45 bar,
–50 psi)
0.0%
Appendix
B.
July 7, 2014, 715003736IVD Rev. C
Signal connections
Parameter Setting at
2.000 Volts (Maximum)
2 mL/min
124,106 kPa (1241 bar,
18,000 psi)
100.0%