Run LED (sample manager)
The run LED, on the right-hand side of the power LED on the sample
manager's front panel, indicates the run status. A steady-green run LED
indicates that injections are being run.
Status LED indications:
LED mode and color
Unlit
Steady green
Flashing green
Flashing red
Steady red

Enabling the leak sensors

When you power-on the system, the leak sensors default to disabled
Rule:
status unless previously enabled.
To enable the leak sensors:
1.
In the ACQUITY UPLC Console, select Control > Leak Sensors.
Description
Binary solvent manager and sample manager –
Indicates the device is currently idle.
• Binary solvent manager – Indicates solvent is
flowing.
• Sample manager – Indicates the sample
manager is operating normally, attempting to
complete any outstanding samples or
diagnostic function requests. When sample
and diagnostic function requests are finished,
the LED reverts to the unlit mode.
Binary solvent manager and sample manager –
Indicates that the module is initializing.
Indicates that an error stopped the instrument or
device. Refer to the ACQUITY UPLC Console for
information regarding the error.
Indicates a module's failure that prevents its
further operation. Power-off the module, and
then power-on. If the LED is still steady red,
contact your Waters service representative.