C External Connections
Binary solvent manager analog-out/event-in connections: (Continued)
Signal connection
0−2V Analog 1 Out
Gradient In
Stop Flow In
Switch 2 Out
Switch 3 Out
0−2V Analog 2 Out
76
July 7, 2014, 715003736IVD Rev. C
Description
Outputs the analog signal to a device such as an
integrator or strip-chart recorder. Via the
MassLynx software, you can select one of the
following signals as the chart out signal:
• Flow rate
• System pressure
• Composition (%A, %B)
Initiates the pumps to begin gradient operation
by contact closure input or 0-volt input.
Allows you to stop the flow from the binary
solvent manager when an error condition or
hardware failure occurs on another system
module.
Used to send time-based contact closure signals
to external devices.
Used to send time-based contact closure signals
to external devices.
Outputs the analog signal to a device such as an
integrator or strip-chart recorder. Via the
MassLynx software, you can select one of the
following signals as the chart out signal:
• Flow rate
• System pressure
• Composition (%A, %B)