Additional functions in the binary solvent manager control panel:
Control panel function
Start up system
Prime solvents
Prime seal wash
Wash plungers
Launch ACQUITY UPLC
Console
Reset BSM
Help
Description
Brings the system to operational conditions
after an extended idle period or when
switching to different solvents.
See "Starting up the system" in the Binary
Solvent Manager Operator's Overview and
Maintenance Information.
Displays the Prime Solvents dialog box.
See "Priming the binary solvent manager" in
the Binary Solvent Manager Operator's
Overview and Maintenance Information.
Starts priming the seal wash.
See "Priming the seal wash system" in the
Binary Solvent Manager Operator's Overview
and Maintenance Information.
Initiates the plunger wash sequence, which
fills and then slowly empties the primary and
accumulator chambers (with the current
solvent composition) while performing a high
speed/volume seal wash. This action helps
prevent the buildup of precipitates on the
pump plungers which can damage the high
pressure seals.
Launches the ACQUITY UPLC Console.
Resets the binary solvent manager after an
error condition.
Displays the ACQUITY UPLC Console
online Help.
July 7, 2014, 715003736IVD Rev. C