Figure 2-10 Waveform display interface of CR Mode

Constant Power (CP) Mode

In CP mode, the electronic load will sink a constant power. If the input voltage
rises, the input current draw will be decreased to maintain a constant power
sink following the equation (P= V*I), as shown in Figure 2-11.
Figure 2-11 Voltage-Current Relationship Schema under CP Mode
Operating Steps
1. Turn off the instrument, as shown in Figure 2-2, connect the DUT
and the channel input terminals on the front panel of the load.
CAUTION
While making a connection, the positive polarity of the load should
be connected to the (+) terminal of the channel output, and the
22