CAUTION
While making a connection, the positive polarity of the load should be
connected to the (+) terminal of the channel output, and the negative
polarity of the load to the (-) terminal of the channel output. Incorrect
polarity may cause damage to the instrument or the DUT.
2. Press CC to enter the main interface of CC mode, as shown in Figure
2-3.
3. Set CC mode current range (5 A or 30 A) and voltage mode (36 V or
150 V)
Note: The lower ranges provide better resolution and accuracy at low
current settings.
4. Set the current input value. This is the amount of current that the load
will attempt to draw when the output is enabled.
5. Set the rising slew rate and the falling slew rate in CC mode. The
default unit for the slew rate is A/us.
6. Press On/Off to turn on the channel input. At this time, the actual input
voltage, current, resistance and power will be displayed on the main
interface.
Note: The load will begin to sink current only if the input voltage is
greater than the conduction voltage of the system (default value
is 0 V).
16
Figure 2-3 CC Mode Main Interface