Teledyne T3EL15030 P Series User Manual page 50

Programmable dc electronic load
Table of Contents

Advertisement

Set Function
Choose CC, CV, CR, or CP mode
Set range
Current range: 5 A or 30 A
Voltage range: 36 V or 150 V
Set Step
Here, you set the number of steps that are required for the list.
Set Count
The Count is the number of times the list will be executed before the
end of the test. The Count has a range from 0-255. When Count is 0,
the SDL will display "infinite" and will run each step in the list until it is
disabled or stopped manually.
Set parameter of list
The parameter list of List mode includes step, sink value of each step,
running time of each step, slew rate, readback current, readback
voltage, readback resistor, readback power. The arrow in the right of
parameter list indicates that the current interface can not display
parameters of all steps. At this time, the user can edit all parameters of
each step by using the direction keys/arrows to move the focus to
different areas of the list.
3. Turn on the input channel
Press On/Off to turn on the channel input. At this time, the actual
discharge voltage, current, time and capacity will be displayed on the
main interface. The parameter status can be displayed in real time in
47