The parameters for the OCPT test mode include range, OCP_V, I_Step,
I_Start, I_End, I_MIN, I_MAX, Delay. The interface menu can be divided
into three pages.
Set range
Current range: 5 A or 30 A
Voltage range: 36 V or 150 V
Set OCP_V
The protection voltage of the OCPT test function. The default unit for
the value is V.
Set I_Step
The step current in OCPT test function. The default unit for I_Step is
Ampere (A).
Set I_Start
The sink current when the load starts the OCPT test sequence. The
default unit for I_Start is Ampere (A).
Set I_End
The sink current when the load stops the OCPT test sequence. The
default unit for I_End is Ampere (A).
Set I_MIN
The minimum value of the protection current in OCPT test function. The
default unit for I_End is Ampere (A).
Set I_MAX
The maximum value of the protection current in OCPT test sequence.
The default unit for I_End is Ampere (A).
Set Delay
The time interval during which the current moves from step-to-step in
the OCPT test sequence. The default unit for Delay is s.
3. Turn on the input channel
Press On/Off to turn on the channel input. At this time, the actual input
voltage, current, resistance and power will be displayed on the main
interface. If the OCPT test passes, the interface of the load will pop-up a
message box with the words "Test complete". Then the load will disable
36