(1) Set operating mode
There are five modes that can be selected: CC, CV, CP, CR, or LED.
(2) Set the current and voltage ranges: Current
range: 5 A or 30 A
Voltage range: 36 V or 150 V
(3) Set pause Status
Users can pause auto test if needed. Simply highlight/select the cell
labelled pause using the keypad arrows on the front panel and rotate the
SDL control knob to enable change the entry value. Then the load will
pause at this step when running this auto test step. Auto test can recover
after receiving the next trigger.
(4) Set short Status
Users can enter a short circuit step if needed. Similar to pause, use the
keypad arrows to select the Short cell and rotate the knob to enable it.
Then the load will short the circuit at this step when running an auto test.
(5) Set Loading Time(Ton)
This is the time duration of each step and its range is 0.01-100 seconds (s).
(6) Set unLoading Time(Toff)
The time interval in adjacent step and its setting range is 0.01-100S
(7) Set Delay Time(Tdly)
The time is between begin test to sample the current or voltage value of
the load. Its setting range is 0.01-Ton s.
(8) Set minimum
The minimum sink value. Units depend on the current mode in use.
� In CC mode, its default unit is V
� In CV mode, its default unit is A
� In CP mode, its default unit is W
� In CR mode, its default unit is Ω
(9) Set maximum
The maximum sink value Units depend on the current mode in use.
� In CC mode, its default unit is V
� In CV mode, its default unit is A
� In CP mode, its default unit is W
� In CR mode, its default unit is Ω
50