Proline Cubemass C 100 HART
Parameter
Measuring mode output
Assign frequency output
Damping output
Measuring mode output
Assign pulse output
Measuring mode output
Operating mode totalizer
Endress+Hauser
Description
Select measuring mode for output.
Select process variable for frequency output.
Set reaction time for output signal to
fluctuations in the measured value.
Select measuring mode for output.
Select process variable for pulse output.
Select measuring mode for output.
Select totalizer calculation mode.
Selection / User entry
• Forward flow
• Forward/Reverse flow
• Reverse flow compensation
• Off
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
• Density
• Reference density
• Concentration
• Dynamic viscosity
• Kinematic viscosity
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Carrier pipe temperature
• Electronic temperature
• Oscillation frequency 0
• Oscillation frequency 1
• Frequency fluctuation 0
• Frequency fluctuation 1
• Oscillation amplitude 0
• Oscillation amplitude 1
• Oscillation damping 0
• Oscillation damping 1
• Tube damping fluctuation 0
• Tube damping fluctuation 1
• Signal asymmetry
• Exciter current 0
• Exciter current 1
0 to 999.9 s
• Forward flow
• Forward/Reverse flow
• Reverse flow
• Reverse flow compensation
• Off
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
• Forward flow
• Forward/Reverse flow
• Reverse flow
• Reverse flow compensation
• Net flow total
• Forward flow total
• Reverse flow total
Commissioning
Factory setting
Forward flow
Off
1 s
Forward flow
Off
Forward flow
Net flow total
67
Need help?
Do you have a question about the Proline Cubemass C 100 HART and is the answer not in the manual?
Questions and answers