Proline Cubemass C 100 HART
Parameter
Burst mode 1 to n
Burst command 1 to n
Burst variable 0
Burst variable 1
Burst variable 2
Burst variable 3
Burst variable 4
Burst variable 5
Burst variable 6
Burst variable 7
Burst trigger mode
Endress+Hauser
Parameter overview with brief description
Description
Activate the HART burst mode for burst
message X.
Select the HART command that is sent to the
HART master.
For HART command 9 and 33: select the
HART device variable or the process variable.
For HART command 9 and 33: select the
HART device variable or the process variable.
For HART command 9 and 33: select the
HART device variable or the process variable.
For HART command 9 and 33: select the
HART device variable or the process variable.
For HART command 9: select the HART
device variable or the process variable.
For HART command 9: select the HART
device variable or the process variable.
For HART command 9: select the HART
device variable or the process variable.
For HART command 9: select the HART
device variable or the process variable.
Select the event that triggers burst message
X.
Min. update period
Max. update period
Selection / User entry
• Off
• On
• Command 1
• Command 2
• Command 3
• Command 9
• Command 33
• Command 48
• Mass flow
• Volume flow
• Corrected volume flow
*
• Target mass flow
*
• Carrier mass flow
• Density
• Reference density
*
• Concentration
• Temperature
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Sensor integrity
• Pressure
• HART input
• Percent of range
• Measured current
• Primary variable (PV)
• Secondary variable (SV)
• Tertiary variable (TV)
• Quaternary variable (QV)
• Not used
See the Burst variable 0
parameter.
See the Burst variable 0
parameter.
See the Burst variable 0
parameter.
See the Burst variable 0
parameter.
See the Burst variable 0
parameter.
See the Burst variable 0
parameter.
See the Burst variable 0
parameter.
• Continuous
• Window
• Rising
• Falling
• On change
System integration
→ 52
→ 52
Factory setting
Off
Command 2
Volume flow
Not used
Not used
Not used
Not used
Not used
Not used
Not used
Continuous
51
Need help?
Do you have a question about the Proline Cubemass C 100 HART and is the answer not in the manual?