Run Preparation
ENTERING PROGRAM 0
RUNNING PROGRAM 0 WITH A SECOND PROGRAM
2-10
To use the delayed start program function, a special program (called
program "0") is created, followed by a second program that contains
the run conditions for the actual run. Follow the instructions below.
The first program —program 0— is used to delay the start of the run
until the required time. Enter the settings as follows.
RTQC
1. Press
then
blinking zero appears above the
for program 0 are displayed.
2. Use the parameter keys and keypad to enter time and temperature
settings as required. You cannot change the speed setting, which
is set at zero. (If you try to change it, the instrument will signal
an error.)
3. Press
. The SAVE message blinks to indicate that
ENTER/RECALL
you can save the program settings.
U7XA
4. Press
to save program 0.
5. To exit from the program mode, press
To run the delayed start program function, do the following.
RTQC
1. Press
then
blinking zero appears above the
the program 0 settings.
2. Use the keypad to select a second program from the stored
programs. The selected program settings appear in the display
and the program number appears in the
.
ENTER/RECALL
3. To begin the delayed start program function, press
UV7TV
and
.
The word "delay" appears in the SPEED display and the second
program number you selected appears in the
on the keypad to access program 0. A
RTQC
key. The existing settings
9A
twice.
on the keypad to access program 0. A
RTQC
key, and the display shows
RTQC
RTQC
display. Press
ENTER/RECALL
display.