Pulse Function; Programmed Run; Program Library - Beckman Coulter Avanti J-15 Series Instructions For Use Manual

Table of Contents

Advertisement

Pulse Function

NOTE
The Time and Accel/Decel settings are overridden by the Pulse function.
The
function eliminates the need to press the
Pulse
Press and hold the
continues to spin as long as the
minutes and seconds.
When the
START
using maximum deceleration.
The centrifuge memory retains the parameters of the last run performed before the
was activated. At the end of a
run parameters are displayed.

Programmed Run

The instrument's internal memory can store up to 99 programs, which can be recalled by selecting
the program number in the
prevent them from being changed. Saved programs are retained in memory even if the power is
turned off.
Figure 2.9 Program Library
1
Turn the power switch on (
PN B80286AC
key. The rotor accelerates at maximum rate to the set speed and
START
key remains pressed. The elapsed time is displayed in
START
key is released, the time stops accumulating and the rotor decelerates to 0 RPM,
run, after the centrifuge door is opened and closed, the previous
Pulse

Program Library

).
I
and
START
STOP
. Programs can be protected with a PIN number to
1. SEARCH — Select to find a program by program
number
2. PROGRAMS — Displays the program number,
speed, time, and temperature (J-15R only)
3. Scroll arrows — Select the up or down arrows
to scroll through the program list
4. Shaded Area - Select anywhere in this area to
return to the Home screen
Operation
Programmed Run
keys for short duration runs.
function
Pulse
2
2-11