Tools Required; Special Tools, Supplies And Equipment Used For Maintenance - GE Voluson E Series Service Manual

Hide thumbs Also See for Voluson E Series:
Table of Contents

Advertisement

Care and Maintenance
Item
Measurement Accuracy
Checks
Probe/Phantom Checks
Check gain changes, vary the focus and check
reaction on screen.
NOTE! The use of a Phantom is not required during
Preventive Maintenance. Customer may use it as part
of their Quality Assurance Program tests.
Functional Checks of all
probes
Section 10.4.2 on
page 10-5

10.3 Tools required

10.3.1 Special Tools, Supplies and Equipment used for Maintenance

Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer tool compliant
Note
with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551 or IEC 62353 or other relevant national regulation.
Table 10-2 overview of supplies and equipment used for maintenance
Tools
Digital Volt Meter (DVM)
Anti Static Kit
Anti Static Vacuum Cleaner
QIQ Phantom
CD-RW Media
DVD+RW Disc Media blank
B/W Printer Cleaning Sheet
Color Printer Cleaning Sheet
Disposable Cloves
10-4
Service at indicated Time
minimum 5% accuracy, 3.5 digit and 200 Ohm range required
including anti-static mat, wrist strap, cables and conductive ground cord
NOTE! The use of a Phantom is not required during Preventive Maintenance. Customer
may use it as part of their Quality Assurance Program tests.
(minimum quad speed)
blank 4,7GB DVD+RW disc
see printer user manual for requirements
see printer user manual for requirements
D W M A
Also after corrective maintenance or
as required by your facilities QA
program.
Also after corrective maintenance or
as required by your facilities QA
program.
Also after corrective maintenance or
as required by your facilities QA
program.
Comments
Voluson E-Series Service Manual
Notes
5539550APB Revision 6