4.5 Software Configuration Checks

4.5.1 Setup

1.
Press the Utilities key on the control console.
2.
In the "Utilities" menu touch the
3.
On the right side of the screen select the desired major group.
Each major group contains different pages to check.
Almost all settings have to be assigned as desired and required by the customer. After completion, return the
Note
equipment to the customer and ensure acceptance.
For further details refer to the Voluson E-Series Basic User Manual.
Note
General Settings
Step
1
General: Check Date and Time setting
2
General: Check that Clinic Name is correct
3
General: Check Language settings
Assign all (further) settings as desired and required by the customer.
Administration
Step
1
Options: Check that all options are set up correct
Connectivity
Step
1
Peripherals: Check the Video Norm standard as
described in
2
Device Setup: Check DICOM, Archive and Network
configuration nodes (e.g., TCP/IP, WLAN, EMAIL, etc.)
Assign all (further) settings as desired and required by the customer.
Backup
not applicable
Imaging Pesets
Assign all (further) settings as desired and required by the customer.
Biopsy
not applicable
Measure
Assign all (further) settings as desired and required by the customer.
Voluson E-Series Service Manual
5539550APB Revision 6
Setup button to invoke the setup desktop on the screen.
Page + Task
Page + Task
Page + Task
Section 3.7.1.6 on page 3-45
.
Functional Checks
Expected Result(s)
Date and Time are correct
Clinic Name (location) is correct
desired System / EUM Language is displayed
Expected Result(s)
D = Demo , I = Inactive , P = Permanent
Expected Result(s)
setting assigned as required for the country
settings assigned as desired and required by the
customer
4-23