Power Off / Shutdown - GE Voluson E Series Service Manual

Hide thumbs Also See for Voluson E Series:
Table of Contents

Advertisement

Functional Checks

4.2.2 Power Off / Shutdown

After turning off a system, wait at least 10 seconds before turning it on again. The system may not be able to
Note
boot if power is recycled too quickly.
1.
If not already in read mode, freeze the image.
2.
Press the ON/OFF Standby button on the control console. Following dialog appears.
Shutdown . The system performs an automatic full shutdown sequence.
3.
Select
4.
Switch OFF the circuit breaker at the rear of the system.
A full shutdown is also performed when pressing the ON/OFF standby button on the control console twice.
Note
The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the ON/OFF
Note
standby button. So the auxiliary equipment need not to be switched ON/OFF separately.
Warning
Disconnection of the main power cable is necessary!
5.
After complete power down, unscrew the 2 screws and remove the pull-out protection to disconnect the
main power cable from the system or unplug it from the AC wall outlet socket.
6.
Press on the brakes to block the front caster wheels.
7.
Disconnect probes. (Turn the probe locking handle counterclockwise and then pull the connector
straight out of the probe port.)
Caution
Do not disconnect a probe while running (Live Scan "Write" mode)! A software error may occur. In this case
switch the system OFF (perform a reset).
4-4
Figure 4-3 Shutdown dialog
Figure 4-4 Circuit Breaker at rear of system
1 circuit breaker
2 fuses (2x T10A H/250V)
3 connector for main power cable
Voluson E-Series Service Manual
5539550APB Revision 6