8.17.1 Mounting/Replacing a Secondary "Patient" Monitor

A Secondary Monitor is NOT intended for diagnostic use. It is an additional device used to allow the patient
Note
to watch the proceedings.
8.17.1.1 Preparing the Isolation Transformer
Caution
For changing the input/output voltage or fuses the isolation transformer should not be connected. The wrong
fuses and position of the input/output voltage selector may cause major damage on connected peripherals.
1.
Before using the Isolation Transformer you must check the input/output voltage settings to meet the
ratings of the line power available in your location or country.
2.
For changing the input/output voltage, remove the label by means of a small socket wrench.
3.
Switch both voltage selectors (input as well as output voltage) to 115V or 230V. Consider that changing
the input/output voltage also requires to change the fuses!
4.
For changing the fuses, open the fuse protection at the power inlet block with a small screwdriver,
remove the fuse holder and exchange fuses.
Figure 8-36 exchange
5.
Assure that the connected loads can be operated with the chosen voltage.
Voluson E-Series Service Manual
5539550APB Revision 6
Figure 8-34 Isolation Transformer kit for Secondary Monitor
Figure 8-35 change input/output voltage
115V (100-130V) -> 4AT
230V (220-240V) -> 2AT
fuses
Replacement Procedures
8-29