Site Preparation

2.2.3 Desirable Features

door is at least 92 cm (3 ft. / 36 in.) wide
circuit breaker for dedicated power outlet is easily accessible
sink with hot and cold water
receptacle for bio–hazardous waste, like used probe sheaths
emergency oxygen supply
storage for linens and equipment
nearby waiting room, lavatory, and dressing room
dual level lighting (bright and dim)
lockable cabinet for software and manuals

2.2.4 Network Setup Requirements

2.2.4.1 Stand-alone System (without Network Connection)
None
2.2.4.2 System connected to Hospital's Network
Supported networks:
Ethernet network connection
Wireless LAN (option)
2.2.4.3 Purpose of the DICOM Network Function
® 1
DICOM
(Digital Imaging and Communications in Medicine) services provide the operator with clinically
useful features for moving images and patient information over a hospital network. Examples of DICOM
services include the transfer of images to workstations for viewing or transferring images to remote printers.
As an added benefit, transferring images in this manner frees up the on-board monitor and peripherals,
enabling viewing to be done while scanning continues. With DICOM, images can be archived, stored, and
retrieved faster, easier, and at a lower cost.
2.2.4.4 DICOM Option Pre-installation Requirements
To configure the Voluson E-Series ultrasound system to work with other network connections, the network
administrator must provide some necessary information.
Figure 3-64 on page 3-73
Use the
Voluson E-Series Details:
Routing Information:
DICOM Application
Information:
For further details refer to the Voluson E-Series Basic User Manual.
Note
1 DICOM is the registered trademark of the National Electrical Manufacturers Association for its standards publications relating to digital communications
of medical information.
2-8
to record required information that must include:
DICOM network details for the Voluson E-Series system, including the
host name, local port, IP address, AE title and net mask.
IP addresses for default gateway and other routers in use at site.
Details of DICOM devices in use at the site, including the DICOM host
name, AE title, DICOM port number and IP addresses.