3.5.9 Connecting a Secondary "Patient" Monitor

Caution
A Secondary "Patient" Monitor MUST NEVER be connected to the Voluson E-Series systems mains supply
directly! Always connect it to an appropriate Isolation Transformer (see
Caution
The secondary monitor is the only item to be connected to the isolation transformer.
A Secondary Monitor is NOT intended for diagnostic use. It is an additional device used to allow the patient
Note
to watch the proceedings.
Take your time to think about the best position of the monitor in your facilities. Patients should be able to
Note
view the monitor easily and without having to bend or turn around.
1.
Power Off/Shutdown the system as described in
2.
Connect the Secondary Monitor according to connection scheme, see:
3.
After physical connection to the Voluson E-Series system check, and if necessary change the Ext.
Monitor Output setting, see:
Connection is always the same (no differences between PC-Motherboard version of the Voluson E-Series
Note
system).
Voluson E-Series Service Manual
5539550APB Revision 6
Section 4.2.2 on page 4-4
Section 3.7.1.7 on page 3-45
Setup Instructions
Section 9.10.3 on page 9-42
.
Figure 3-22 on page 3-30
.
).
.
3-29