How To Activate User Login Authentication; How To Perform Full Disk Encryption - GE Voluson E Series Service Manual

Hide thumbs Also See for Voluson E Series:
Table of Contents

Advertisement

Setup Instructions

3.7.1.12 How to activate User Login Authentication

1.
Invoke System Setup as described in
2.
On the right side of the screen select
3.
Check mark
a.
Click
b.
Enter the
c.
Enter a
Note
A valid password must be at least 6 characters long and has a maximum length of 80 characters. The
password must contain at least 2 non-letter characters, 0...9 or !@#$%^*().
d.
Select a
e.
Confirm with
4.
Close the Service page with
The system login window appears as soon as the system is started or a current user logs off. The correct
User Name and Password has to be entered to login. If the correct password/user name combination is
entered, it is also possible to change the password. The
restricted access.
For further details refer to the Voluson E-Series Basic User Manual.
Note

3.7.1.13 How to perform Full Disk Encryption

Disk Encryption enables an encryption to secure all Patient Data, Patient Images and corresponding
measurements.
Caution
Without the Encryption Password or Recovery Key it will not be possible to access the Ultrasound device
including emergency mode, scanning, patient information, images, or local archive. GE has no access to this
information or the ability to undo encryption in the event that the Encryption Password and Recovery Key are
lost. The security and safety of the Encryption Password and Recovery Key is solely the user's responsibility.
Note
Disk Encryption is an option and might not be available in all countries. If the option is not set, the tab Disk
Encryption is not available in System Setup - Administration.
Content in this section
3.7.1.13.1 Disk Encryption Workflow - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-47
3.7.1.13.2 Unlock an encrypted Disk - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-47
3.7.1.13.3 Decrypt Disk - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-48
3-46
Section 3.7.1.1 on page 3-43
General Settings and then click the Authentication tab.
Activate User Login .
Figure 3-36 System Setup - Administration - Authentication
Add User . A dialog appears.
User Name , the First Name and the Last Name .
Password and Confirm Password .
Group the user belong to.
OK to save User Login Authentication.
Save&Exit and restart the system.
.
Emergency button enables emergency mode with
Voluson E-Series Service Manual
5539550APB Revision 6

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voluson e6Voluson e8Bt15Bt16Bt17Bt18

Table of Contents