Isolated Patient Lead (Sink) Leakage-Isolation Test - GE Voluson E Series Service Manual

Hide thumbs Also See for Voluson E Series:
Table of Contents

Advertisement

Care and Maintenance

10.6.8 Isolated Patient Lead (Sink) Leakage-Isolation Test

Select the ALL position on the LEAD selector since the test is performed with mains applied to all ECG leads
at the same time.
Caution
Line voltage is applied to the ECG leads during this test. To avoid possible electric shock hazard, the system
being tested must not be touched by patients, users or anyone while the ISO TEST switch is depressed.
10.6.8.1 Isolated Lead (Sink) Leakage Test Record
Table 10-16
shows a typical format for recording the isolated patient lead sink leakage current.
Measurements should be recorded for full lead combination under each set of test conditions specified in:
Table 10-11 on page 10-11
Record all data and keep the record of the results with other hard copies of maintenance data.
1.
Connect the Safety analyzer to an AC wall outlet.
2.
Plug the equipment under test into the receptacle on the panel of the meter.
3.
Connect the ECG cable to the Voluson E-Series system and the patient leads to the analyzer.
4.
Select the "Patient Lead Leakage" function on the meter.
5.
Test with closed ground with the Voluson E-Series system on and off.
For more information, refer to the safety analyzer's user manual that will be used to perform the test.
Note
Unit under Test:
System Power
on
off
on
off
Values in italics font are given as examples only.
Note
10-18
Table 10-16 Typical Data Format for recording Isolated Patient Lead (sink) Leakage
Test Conditions
Grounding/PE
closed
closed
closed (reversed polarity)
closed (reversed polarity)
Date of Test:
Patient Lead
RA+LA+LL
Limit μA
50
50
50
50
Voluson E-Series Service Manual