The following screen appears.
10.
Select the
11.
Press Alt + D to export system data and the needed logs. Operation see
System Data (by pressing the ALT + D key)" on page 7-10
The standalone recorded "Auto Tester" file makes only sense with the Alt + D log! To analyze the workflow
Note
the exact date and time of occurrence needs to be documented!
12.
If desired, save the script file also on DVD/CD. Therefore:
-
Insert an empty DVD/CD+R/RW in the Drive and select the
-
Enter a Filename.
-
After the DVD/CD write is finished click the
13.
Close the "Auto Tester" program.

7.7.1 Limitation of the Auto Tester

The following information will not be recorded. Depending on the moment the Auto Tester is activated, this
must be provided by the customer or the field engineer.
Which probe is in use and which probes are connected?
Which Mode is activated?
Which peripherals are connected (Dicom, Printer, etc..)?
Collected information from all steps above + exported system data and needed logs (Alt + D) can be sent to
the Make Center. With this information the Make Center can see how the customer is using the system and
reproduce potential failures.
Providing all information at once will help the Make Center find the root cause and speeds up finding a
Note
solution for the customer.
Voluson E-Series Service Manual
5539550APB Revision 6
Figure 7-25 Autotester finished
"Save" icon. The file will be saved in D:\Export\AutotesterScripts\*.*.
Diagnostics/Troubleshooting
Section 7.4.2.1 "Export
.
"Burn on CD" icon.
OK button.
7-19