7.5.2 How the Customer enables/disables Disruptive Mode and VCO

1.
If not already in read mode, Freeze the image.
2.
Move the cursor to the InSite link (
3.
Press the right trackball key (= right-click).
Connect Clinical Lifeline (see:
4.
Select
for the application OLC to quickly assist the customer.
Figure 7-12 Connect Clinical Lifeline
Cancel respectively Disconnect to turn OFF "Disruptive Mode" and "VCO".
5.
Select
Visual information about GE remote status is shown in the status area on the right bottom of the screen. At
Note
EC330 (BT18) systems the trackball functions are displayed in the trackball status area of the screen.
Gray = Idle State
Yellow = Access
pending
Red = Disrupted
State
Gray + envelope
If a GE Service Technician requests Remote Access Permission
If a GE Service technician requests remote access to your Voluson E-Series system, following message
appears on the screen.
To enable Disruptive Mode click
If the customer does not wish to have diagnostics running at the time of the request, select Cancel. A
Note
message is sent back to the OLC or FE that permission to change Disruptive Status is denied.
Voluson E-Series Service Manual
5539550APB Revision 6
GE icon at the right bottom of the display screen).
Figure 7-12 below
Remote access is inactive.
EC330 only: Move the cursor over the
Contact :
Section 7.2 "Request for Service (RFS)" on page 7-5
Configuration :
Menu : Displays the Remote Access (Disruptive Mode) menu
Remote Access connection is active, but Disruptive Mode and Virtual Console
Observation (VCO) is not yet enabled.
Remote Access is active. All processes [UL_VNC and UL_CSD] are active. In this
state the Voluson E-Series system should NOT be used clinically.
New Software Update Package is available for download and installation.
Operation see
Section 8.2.3 on page 8-10
Figure 7-13 Insite User Session Message
OK .
). This activates "Disruptive Mode" and "VCO"
GE icon to show trackball functions
Section 3.13.8 "InSite Configuration" on page 3-70
.
Diagnostics/Troubleshooting
7-13