8.3 Software and Functional Checks after Installation/Upgrade procedure

1.
Press the Utilities key on the control console.
2.
In the "Utilities" menu touch the
3.
On the right side of the screen select
4.
Verify the correct settings of the
the settings according to the printout.
5.
Click the
1
Ultrasound Application Software version
2
Backup File Name
6.
Check the Application Software version.
7.
Check that the Tune version of the Application presets match the Application Software version.
It is neither required nor advisable to reload a previously stored "Full Backup" after a software upgrade that
Note
was performed by means of the FMI from DVD button! If the Tune version does not match the Application
Software version, a warning message appears whenever booting up the system. In this case, it is essential
to load the proper Application Settings (image presets), adapted for the installed software version. Refer to:
Section 5.14.3.14 "Load Application Settings" on page 5-52
8.
Perform basic functional checks to ensure system is functioning normally.
9.
Check Service Connectivity; if required, perform InSite Checkout. i.e., Service platform has to be
configured properly; see
8-14
Setup button to invoke the setup desktop on the screen.
Options page; see:
System Info tab.
Figure 8-19 System Setup - Administration - System Info page
Section 7.6.5 "Configuration" on page 7-15
Administration and then click the Options tab.
Figure 8-2 on page 8-5
3
Tune version of Application presets
.
. If necessary, customize
.
Voluson E-Series Service Manual