3.
Select the desired file and then click
Load procedure starts immediately including a reboot of the system.
If the ID of the Application Setting is not valid for the currently installed Application Software version, a
Note
warning message appears during boot up sequence.
4.
If warning message is displayed, confirm it with
(perform loading procedure as described in steps above).

5.15 Boot Screen Functions

Following LINUX supported functions are available as soon as the "Boot Screen" appears:
Content in this section

5.15.1 Voluson - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-53

5.15.2 GE-Service - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-53

5.15.3 Rollback - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-53

5.15.4 Memtest (Memory Check) - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5-54
After 3 sec. without pressing any key, the system will boot-up in windows (= Voluson item). If you missed
selection, retry again with Ctrl + Alt + Del.
5.15.1 Voluson
The System will boot-up in windows. The Ultrasound Application is started. For details see
"During a normal boot, you may observe" on page 3-11
5.15.2 GE-Service
This function MUST NOT be used by the customer!
5.15.3 Rollback
This function offers the possibility to simply restore the previously used system configuration (rollback), which
was stored on R:\ during "FMI from DVD".
1.
Turn system OFF and then back ON.
2.
As soon as the "Boot Screen" appears (see:
down) key on the keyboard until the Rollback item is highlighted, then press Enter.
Voluson E-Series Service Manual
5539550APB Revision 6
Ok .
Figure 5-29 Boot screen
Figure 5-30 on page 5-53
Components and Functions (Theory)
OK and then load appropriate Application Settings
Voluson
GE-Service
Rollback
Memtest (Memory Check)
.
), press the [PgDn] (Arrow
Section 3.4.1.1