Caution
Follow the Care Card instructions supplied with each probe (inside the probe boxes) for disinfectants and
gels that are compatible with the surface material of the probes.
For the latest list of compatible cleaning solutions and disinfectants refer to: http://www.gehealthcare.com/
Note
transducers.
Please be aware of the
sensitive probe head. TAKE
EXTREME CARE!
10.4.5.4 Disinfection by means of the RIC-Holder
Especially for Real-time 4D endocavity probes (RIC), it is necessary to take extreme care when transporting
the system with the probe attached, or during the disinfection process. Inadequate handling may lead to
dead elements, shocked head mechanics, etc.
The RIC-Holder (especially developed for RIC Real-time 4D endocavity probes) guarantees that the
sensitive probe head does not hit the bottom of the container during the disinfection procedure.
Instructions are supplied with each RIC-Holder (KTZ225469).
Note

10.5 Using a Phantom

The use of a Phantom is not required during Preventive Maintenance. Customer may use it as part of their
Quality Assurance Program tests.
Voluson E-Series Service Manual
5539550APB Revision 6
NEVER place or store a probe
on its scan head!
Care and Maintenance
When disinfecting a probe,
ensure that there is sufficient
space between the probe and
the container bottom!
10-9