WARNING
Risk of improper operation.
Use of accessories, transducers and cables
other than those specified or provided by the
manufacturer of this equipment could result in
increased electromagnetic emissions or
decreased electromagnetic immunity of this
equipment and result in improper operation.
Use only accessories, transducers and cables
specified or provided by the manufacturer of
this equipment.
Not for use in areas of explosion hazard
or oxygen-enriched atmospheres
WARNING
Risk of fire
The medical device is not approved for use in
areas where combustible or explosive gas
mixtures are likely to occur or in oxygen-
enriched atmospheres.

Connected devices

WARNING
Risk of electric shock and of device failure
Any connected devices or device combina-
tions not complying with the requirements
mentioned in these instructions for use may
compromise the correct functioning of the
medical device. Before operating any combi-
nation of devices, refer to and strictly comply
with the instructions for use for all connected
devices and device combinations.
Instructions for use - Infinity
®
Acute Care System - Infinity
For your safety and that of your patients
Safe connection with other electrical
equipment
WARNING
The leakage current increases when multiple
medical devices are connected to a patient.
Make sure that the galvanic isolation of each
device is suitable for the intended application.
Connect only equipment that is set up and
tested according to IEC standards to the
analog and digital signal inputs and outputs.
Connect only passive USB devices to the
IACS (Infinity Acute Care System) Cockpit.
To protect the patient from possible injury due
to electrical shock, peripheral devices should
only be connected to a monitor within the
same room. The installer or service provider
should verify that the leakage current of the
interconnected system meets the electrical
safety requirements of IEC 60601-1 and
IEC 62353.
CAUTION
Risk of patient injury
Electrical connections to equipment not listed in
these instructions for use must only be made
when approved by each respective manufacturer.

Connection to IT network

The medical device has interfaces to connect to
other devices or components. These interfaces are
only suitable for connecting devices or connecting
to networks that have the physical characteristics
described on page 12 and page 52.
For software-related requirements on the
connection of devices or networks to these
interfaces, observe the relevant documents of
these applications.
If the Cockpit is used to send or distribute alarm
conditions to another device and that system is not
able to be a "distributed alarm system with
confirmed delivery of alarm conditions" as per IEC
60601-1-8:2006 +A1:2012, cl. 6.11.2.2, then the
®
Medical Cockpits
9