Using Transpector Web Ui; Transpector Web Ui And Fabguard - Inficon Transpector MPH Operating Manual

Gas analysis system
Table of Contents

Advertisement

5.5 Using Transpector Web UI

This section provides instruction for common Transpector MPH applications.

5.5.1 Transpector Web UI and FabGuard

FabGuard is the primary software package for the Transpector MPH sensor, but
there are instances when using Transpector Web UI is appropriate.
A FabGuard parameter defines when FabGuard allows Transpector Web UI to
access the Transpector MPH sensor. This parameter has three options:
Never: Sensor Cannot Be Independently Operated
FabGuard will never give control of the Transpector MPH sensor to
Transpector Web UI. This is the Default setting inside of FabGuard.
If Stopped: Only if the Chamber is Stopped
FabGuard will give control of the Transpector MPH sensor only if the chamber
on which the sensor is installed is stopped.
If Stopped or Waiting: Only if the Chamber is Stopped or Waiting for
Acquisition Trigger
FabGuard will give control of the Transpector MPH sensor if the chamber is
stopped or if the chamber is waiting for an acquisition trigger.
CAUTION
FabGuard and Transpector Web UI cannot access the
same sensor at the same time.
Transpector MPH Operating Manual
5 - 10

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Transpector MPH and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents

Save PDF