Figure 8.28 Bin Format Setting - Rigol HDO4000 Series User Manual

Digital oscilloscope
Hide thumbs Also See for HDO4000 Series:
Table of Contents

Advertisement

Triggering the Oscilloscope
MissedAck: triggers when the SDA data is high level during any
acknowledgment of SCL clock position.
Address: the trigger searches for the specified address value. When this event
occurs, the oscilloscope will trigger on the read/write bit. After this trigger
condition is selected:
-
-
-
Data: the trigger searches for the specified data value on the data line (SDA).
When this event occurs, the oscilloscope will trigger on the clock line (SCL)
transition edge of the last bit of data. After this trigger condition is selected, you
can set the length of bytes and data format.
-
-
HDO4000 User Guide
106
Click or tap the drop-down button of Direction to select "Write", "Read", or
"R/W".
This setting is not available when AddrBits is set to "8 Bits".
Click or tap the drop-down button of AddrBits to select the desired
address bits. The available address bits are "7 Bits", "8 Bits", and "10 Bits".
Click or tap the input field of Address, and then use the pop-up numeric
keypad to set the address of I2C trigger. You can also use the corresponding
multipurpose knob to set the value.
Click or tap the input field of Bytes, and then use the pop-up numeric
keypad to set the length of the data. You can also use the corresponding
multipurpose knob to set the value. Its range is from 1 to 5.
Click or tap the input field of Data, and then the "Format" interface is will
displayed. As shown in the figure below, you can select "Bin" or "Hex" data
format.

Figure 8.28 Bin Format Setting

Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hdo4204Hdo4404Hdo4804

Table of Contents