Vertical System
may occur on the zero-cross voltage of the channel, which will affect the
measurement results of the vertical parameters. This series oscilloscope allows you to
set an offset calibration voltage for calibrating the zero point of the corresponding
channel, so as to improve the accuracy of the measurement results.
In "Vertical" menu, click or tap the Up/Down arrow icon at the right side of the input
field of Bias to increase/decrease the bias value or use the corresponding
multipurpose knob to set the value. You can also click or tap the input field to input a
specific value with the displayed numeric keypad.
The range of bias is related to the input impedance and the vertical scale.
TIP
If the zero-cross voltage of the channel has a larger amplitude offset that exceeds the
adjustable null range, please perform self-calibration for the instrument to ensure the
measurement accuracy. For details, please refer to
5.11

To Specify the Skew

When using an oscilloscope for actual measurement, the transmission delay of the
probe cable may bring relatively greater errors (zero offset). This series oscilloscope
allows you to set a delay time for calibrating the zero offset of the corresponding
channel. Zero offset is defined as the offset of the crossing point of the waveform and
trigger level line relative to trigger position, as shown in the figure below.
HDO4000 User Guide
52
SelfCal
.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.