Rigol DS1000D series User Manual

Rigol DS1000D series User Manual

Ds1000e, ds1000d series digital oscilloscopes
Hide thumbs Also See for DS1000D series:

Advertisement

RIGOL
User's Guide
Publication number UGA07107-1110
July 2008
DS1000E, DS1000D Series
Digital Oscilloscopes
DS1102E, DS1052E, DS1102D, DS1052D
© 2008 RIGOL Technologies, Inc.
All Rights Reserved

Advertisement

loading
Need help?

Need help?

Do you have a question about the DS1000D series and is the answer not in the manual?

Questions and answers

Summary of Contents for Rigol DS1000D series

  • Page 1 RIGOL User’s Guide Publication number UGA07107-1110 July 2008 DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D © 2008 RIGOL Technologies, Inc. All Rights Reserved...
  • Page 3  Information in this publication replaces all previously corresponding material.  RIGOL Technologies, Inc. reserves the right to modify or change part of or all  the specifications and pricing policies at company’s sole decision. NOTE: RIGOL is registered trademark of RIGOL Technologies, Inc.
  • Page 4: Safety Notices

    Avoid Circuit or Wire Exposure. Do not touch exposed connections and components when power is on. Do Not Operate With Suspected Failures. If suspected damage occurs with the instrument, have it inspected by qualified service personnel before further operations. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 5 The DS1000E, DS1000D series Digital Oscilloscopes is designed to safely withstand occasional transient overvoltage up to 1000Vpk. Do not use this equipment to measure circuits where transient overvoltage could exceed this level. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 6: Safety Terms And Symbols

    CAUTION indicates that a potential damage to the instrument or other property might occur. Symbols on the Product: These symbols may appear on the Instrument: Hazardous Refer to Protective Grounding Test Voltage Instructions Earth Terminal Terminal Grounding Chassis Terminal © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 7 Oscilloscopes: DS1102E, DS1052E, DS1102D, DS1052D. (With Logic Analyzer) RIGOL DS1000E, DS1000D Series Digital Oscilloscopes provide exceptional waveform viewing and measurements in a compact, lightweight package. The DS1000E, DS1000D series is ideal for production test, field service, research, design, education and training applications involving analog/digital circuits test and troubleshooting, as well as education and training.
  • Page 8  Pop-up menu makes it easy to read and easy to use Built-in Chinese and English help system  Easy-to-use file system supports Chinese & English characters input  © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 9: Table Of Contents

    To Set up the Sampling System .............. 2-61 To Set up the Display System ..............2-65 To Store and Recall ................2-67 To Set up the Utility ................2-75 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 10 Prompting Message ................4-1 Troubleshooting ..................4-3 Chapter 5 : Specifications ..............5-1 Specifications ..................5-2 General Specifications ................5-6 Chapter 6 : Appendix ................. 6-1 Appendix A: Accessories ................6-1 © 2008 RIGOL Technologies, Inc. VIII User’s Guide for DS1000E, DS1000D Series...
  • Page 11 RIGOL Appendix B: Warranty ................6-2 Appendix C: Care and Cleaning ..............6-3 Appendix D: Contact RIGOL ..............6-4 Index ....................6-1 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 13: Chapter 1 : Getting Started

     To use digital leads  To display a signal automatically  To understand the vertical system  To understand the horizontal system  To trigger the oscilloscope  © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 14: The Front Panel And User Interface

     The front panel of DS1000E:  The front panel of DS1000D: Figure 1- 1 Front Panel Figure of DS1000E, DS1000D Series Oscilloscope © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 15 SCALE knobs.  denotes the LEVEL knob.  The name with a drop shadow denotes the menu operating key, such as  WAVEFORM soft key in STORAGE menu. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 16 Running status turned off turned on Channel 1 Digital channels Horizontal Channel coupling Trigger time base div. and vertical div. offset Figure 1- 4 User Interface (Analog and Digital channels) © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 17: To Inspect The Instrument

    Representative. If the shipping container is damaged, or the cushioning materials show signs of stress, please notify the carrier as well as the RIGOL sales office. Keep the shipping materials for the carrier’s inspection. RIGOL offices will arrange for repair or replacement at RIGOL’s option without waiting for claim settlement.
  • Page 18: To Perform A Functional Check

    DS1000D series: 2 channels input + 1 external trigger channel input +16 channels digital input Do the following steps: ① Set the switch on the probe to 10X and connect the probe to Channel 1 on the © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 19 Push the CH2 button to turn on channel 2, repeat steps 2 and 3. NOTE: The signal output from Probe compensator should only be used for probe compensation, not for calibration. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 20: To Compensate Probes

    WARNNING: To avoid electric shock while using the probe, be sure the perfection of the insulated cable, and do not touch the metallic portions of the probe head while it is connected with a voltage source. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 21: Digital Leads (Ds1000D Series)

    Figure 1- 10 Connect the digital leads CAUTION: Use only FC1868, LH1116, TC1100 and LC1150 made by RIGOL for specified DS1000D series. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 22 Test clip Figure 1- 11 4. Test your device with the clip. Figure 1- 12 5. Remember to connect Ground Channel to the DUT’s ground terminal. Figure 1- 13 © 2008 RIGOL Technologies, Inc. 1-10 User’s Guide for DS1000E, DS1000D Series...
  • Page 23: To Display A Signal Automatically

    2. Press AUTO. The oscilloscope may change the current settings to display the signal; and adjusts the vertical and horizontal scaling, the trigger coupling, type, position, slope, level, and mode. © 2008 RIGOL Technologies, Inc. 1-11 User’s Guide for DS1000E, DS1000D Series...
  • Page 24: To Understand The Vertical System

    If the channel is AC coupled, the DC component of the signal is blocked, allow you to use greater sensitivity to display the AC component of the signal. © 2008 RIGOL Technologies, Inc. 1-12 User’s Guide for DS1000E, DS1000D Series...
  • Page 25 Press OFF button to turn off the channel.  Coarse/Fine Shortcut key The Coarse/Fine vertical control can be set by simply pressing the vertical knob. © 2008 RIGOL Technologies, Inc. 1-13 User’s Guide for DS1000E, DS1000D Series...
  • Page 26: To Understand The Horizontal System

    Delayed Scan mode and it is equal to the following menu operations, MENU→Delayed→ON. © 2008 RIGOL Technologies, Inc. 1-14 User’s Guide for DS1000E, DS1000D Series...
  • Page 27: Horizontal Position Control

    Horizontal position control Trig-Offset: In this setting, the trigger position will be changed horizontally when you turn the knob. © 2008 RIGOL Technologies, Inc. 1-15 User’s Guide for DS1000E, DS1000D Series...
  • Page 28: To Trigger The Oscilloscope

    双 Trigger Level to 0 Shortcut key Turn the knob to change trigger level value and press the knob to set trigger level back to 0 as a shortcut key. © 2008 RIGOL Technologies, Inc. 1-16 User’s Guide for DS1000E, DS1000D Series...
  • Page 29 This function helps to view complex signals such as an AM waveform. Press Holdoff button to activate ( ) knob, then turn it to adjust Holdoff time. © 2008 RIGOL Technologies, Inc. 1-17 User’s Guide for DS1000E, DS1000D Series...
  • Page 30 RIGOL © 2008 RIGOL Technologies, Inc. 1-18 User’s Guide for DS1000E, DS1000D Series...
  • Page 31: Chapter 2 : Operating Your Oscilloscope

    To set up utility ( Utility)  To measure automatically ( Measure)  To measure with cursors ( Cursor)  To use run control buttons ( AUTO, RUN/STOP)  © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 32: To Set Up The Vertical System

    100X vertical scale readout correct 500X 1000X Setup digital filter (See table 2-4) Digital filter Go to the next menu page (The followings are the same, no more explanation) © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 33 Defines a 1-2-5 sequence. Volts/Div To change the resolution to small Fine steps between the coarse settings. Turn on the invert function. Invert Restore original display of the waveform. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 34: Channel Coupling

    AC coupling setting Press CH1→Coupling→DC, to set “DC” coupling. It will pass both AC and DC components of the input signal. The waveform is displayed as Figure 2- 4: © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 35 Press CH1→Coupling→GND, to set “GND” coupling, it disconnects the input signal. The screen displays as Figure 2-5: GND coupling setup GND coupling status symbol Figure 2- 5 GND coupling setting © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 36 Turn off the BW limit Press CH1→BW Limit→ON, to set up bandwidth limit to “ON” status. It will reject the frequency component higher than 20MHz. The waveform is displayed as Figure 2- 7: © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 37 RIGOL 20M BW limit Mark of BW limit Figure 2- 7 Turn on the BW limit © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 38: Probe Attenuation Setting

    Probe attenuation Vertical volt/div. Figure 2- 8 Use the 1000:1 attenuation Table 2- 3 Probe setting Probe attenuation factors Corresponding settings 10:1 50:1 100:1 100X 500:1 500X 1000:1 1000X © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 39 It will be helpful to adjust the waveform in smooth steps. Fine adjustment Fine adjustment data Figure 2- 9 Fine configurations Coarse/Fine Shortcut key: To change Coarse/Fine setting, not only by menu but also by pressing vertical knob © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 40: To Invert A Waveform

    Figure 2- 10 and Figure 2- 11 show the changes after inversion. Invert OFF Figure 2- 10 The waveform before inversion Invert ON Figure 2- 11 The waveform after inversion © 2008 RIGOL Technologies, Inc. 2-10 User’s Guide for DS1000E, DS1000D Series...
  • Page 41 Figure 2- 12 The waveform when turning off digital filter Turn on digital filter Mark of digital filter Figure 2- 13 The waveform when turning on digital filter 2-11 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 42 ) knob to set high limit <frequency> Lower limit Turn ( ) knob to set low limit <frequency> Back to higher level menu (The followings are the same, no more explanation) © 2008 RIGOL Technologies, Inc. 2-12 User’s Guide for DS1000E, DS1000D Series...
  • Page 43: Math Functions

    Source A Define CH1 or CH2 as source A Source B Define CH1 or CH2 as source B Invert the MATH waveform. Invert Restore to original waveform display. 2-13 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 44: Using The Fft

    To display FFT waveforms with a large dynamic range, use the dBVrms scale. The dBVrms scale displays component magnitudes using a log scale. © 2008 RIGOL Technologies, Inc. 2-14 User’s Guide for DS1000E, DS1000D Series...
  • Page 45: Selecting An Fft Window

    It’s normally half of the sample rate. This frequency is called the Nyquist frequency. Frequency above the Nyquist frequency will be under sampled, causing a situation known as aliasing. 2-15 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 46: Using Ref

    Select memory location in scope Location External Select memory location out scope Save REF waveform to outer memory Save location Import Go to import menu(see table 2-14) Reset Reset REF waveform © 2008 RIGOL Technologies, Inc. 2-16 User’s Guide for DS1000E, DS1000D Series...
  • Page 47 Import the REF file to internal Import memory Delete Delete file File The figure of import and export as following Figure 2- 21 Import of export the figure 2-17 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 48 Comments Move the cursor up Move the cursor down To delete chosen letter Save Execute the operation The figure of export as following. Figure 2- 23 Figure export © 2008 RIGOL Technologies, Inc. 2-18 User’s Guide for DS1000E, DS1000D Series...
  • Page 49 Set up new file in Path and File. (Folder) Set up new folder in directory. Delete Delete file(Folder) File(Folder) The figure of Save as following: Figure 2- 25 Save the figure 2-19 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 50 Execute the operation The figure of key in as following File name key in Spell key in Character window Switch Capital on/off Switch Chinese/English Figure 2- 27 Chinese Input © 2008 RIGOL Technologies, Inc. 2-20 User’s Guide for DS1000E, DS1000D Series...
  • Page 51 Switch among Path, Directory and Explorer Directory File File Import the REF file into internal Import memory The figure of import as following. Figure 2- 29 Figure import 2-21 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 52 4. Press soft button No.2 to select the save location of REF waveform. 5. Press soft button No.3 to save the waveform as REF. NOTE: The reference function is not available in X-Y mode. © 2008 RIGOL Technologies, Inc. 2-22 User’s Guide for DS1000E, DS1000D Series...
  • Page 53: Set Up La Channel (Ds1000D Series)

    (2) Press LA→current and choose digital channel by turning ( ) knob. The chosen channel will display in red color. (3) Turn vertical knob to re-position the channel in screen. 2-23 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 54 Press LA→Threshold, select logic standard or User to define your own threshold voltage. The figure of menu shows as follow. Setup threshold Setup data of threshold by user Figure 2- 33 Set the threshold © 2008 RIGOL Technologies, Inc. 2-24 User’s Guide for DS1000E, DS1000D Series...
  • Page 55 RIGOL Threshold explanation LOGIC STANDARD THRESHOULD VLOTAGE 1.4V CMOS 2.5V -1.3V User -8V to +8V 2-25 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 56 Turn on or off 8 channels together Turn off Display 8 channels in a single screen Size Display 16 channels in a single screen Reset Reset waveform of channel D15-D8 © 2008 RIGOL Technologies, Inc. 2-26 User’s Guide for DS1000E, DS1000D Series...
  • Page 57 8 channels on the screen; Select to view all of the 16 channels on the screen. 4.Reset the logic channels display: Press LA→D7-D0→Reset, or D15-D8→Reset to reset the display of logic channels. 2-27 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 58: Turn On/Off Channels

    Selected Math (purple letter) No indicator NOTE The channel status symbol is displayed at the lower-left of the screen. Pressing LA will turn all the digital channels on/off. © 2008 RIGOL Technologies, Inc. 2-28 User’s Guide for DS1000E, DS1000D Series...
  • Page 59: Set Up Vertical Position And Scale

    4. During the vertical position, a position message is displayed on the left bottom of the screen, in the same color as the corresponding channel. The unit is V (Volts). 2-29 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 60: To Set Up The Horizontal System

    Delayed Scan time base. Horizontal Menu. Press the horizontal MENU button to display the horizontal menu. The settings of this menu are listed in the following table. © 2008 RIGOL Technologies, Inc. 2-30 User’s Guide for DS1000E, DS1000D Series...
  • Page 61 Sa Rate Show system sample rate Trig-offset Reset Adjust to the center ① ② ③ ④ ⑤ Figure 2- 38 Status bar and mark for Horizontal control 2-31 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 62 DC. Time/Div: Horizontal scale. If the waveform acquisition is stopped (using the RUN/STOP button), the Time/Div control expands or compresses the waveform. © 2008 RIGOL Technologies, Inc. 2-32 User’s Guide for DS1000E, DS1000D Series...
  • Page 63: Delayed Scan

    The horizontal knobs control the size and position of the Delayed Scan. The value at bottom of the screen is the main time base and the 2-33 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 64 Notice the changes in the status bar. Delayed Scan Shortcut Key: Delayed Scan function can be activated not only by menu but also by pressing horizontal knob. © 2008 RIGOL Technologies, Inc. 2-34 User’s Guide for DS1000E, DS1000D Series...
  • Page 65 LA Function (DS1000D series)  Automatic Measurements  Cursor Measurements  REF and MATH Operations  Delayed Scan Mode  Vector Display Mode  Horizontal knob  Trigger Controls  2-35 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 66: To Set Up The Trigger System

    Force to create a trigger signal and the function is mainly used in Normal and Single mode MENU: The button that activates the trigger controls menu. Figure 2- 41 Trigger controls © 2008 RIGOL Technologies, Inc. 2-36 User’s Guide for DS1000E, DS1000D Series...
  • Page 67: Trigger Modes

    Alternative: Trigger on non-synchronized signals Pattern: To Trigger through detecting a specified code. Duration: To trigger within a specified time on the conditions of a specified code 2-37 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 68: Settings For Edge Trigger

    Sweep Normal Acquire waveform when trigger occurred. Single When trigger occurs, acquire waveform then stop Set up To go to Set Up menu, see table 2-38 © 2008 RIGOL Technologies, Inc. 2-38 User’s Guide for DS1000E, DS1000D Series...
  • Page 69: Settings For Pulse Width Trigger

    (+Pulse width equal to) When To select pulse condition (-Pulse width less than) (-Pulse width more than) (-Pulse width equal to) Settings Set required pulse width <Width> 2-39 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 70 To go to Set Up menu, see table 2-38 NOTE: The Pulse width adjust range is 20ns ~ 10s. When the condition is met, it will trigger and acquire the waveform. © 2008 RIGOL Technologies, Inc. 2-40 User’s Guide for DS1000E, DS1000D Series...
  • Page 71: Settings For Video Trigger

    Trigger on all lines Line Num Trigger on an specified line Sync Odd field Select to trigger on odd field Even field Select to trigger on even field 2-41 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 72 Acquire waveform when trigger Sweep Normal occurred. Single When trigger occurs, acquire one waveform and then stop To go to set up menu, see table Set Up 2-39 © 2008 RIGOL Technologies, Inc. 2-42 User’s Guide for DS1000E, DS1000D Series...
  • Page 73 Sync Pulses: When Normal Polarity is selected, the trigger always occurs on negative-going sync pulses. If the video signal has positive-going sync pulses, use the inverted Polarity selection. Figure 2- 48 Video Trigger: Line Synchronization 2-43 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 74 RIGOL Figure 2- 49 Video Trigger: Field Synchronization © 2008 RIGOL Technologies, Inc. 2-44 User’s Guide for DS1000E, DS1000D Series...
  • Page 75: Slope Trigger

    Sweep Normal condition is met. When trigger condition is met, acquire Single one waveform and then stop Set Up To go to set up menu. See table 2-38 2-45 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 76 NOTE: Slope time can be set from 20ns to 10s. When a signal meets the trigger condition, scope will execute the acquisition. You can adjust LEVEL A/ LEVEL B or both simultaneous by turning the knob. © 2008 RIGOL Technologies, Inc. 2-46 User’s Guide for DS1000E, DS1000D Series...
  • Page 77: Alternative Trigger

    Set Edge Trigger as the trigger Type Edge type (Rising) Trigger on rising edge Slope (Falling) Trigger on falling edge To go to set up menu. See table Set Up 2-38 2-47 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 78 Figure 2- 54 Table 2- 30 The Alternative menu (Trigger Type: Pulse, Page 2/2) Menu Settings Comments Setting Set the width of the pulse < > pulse width Set Up To go to set up menu. See table 2-38 © 2008 RIGOL Technologies, Inc. 2-48 User’s Guide for DS1000E, DS1000D Series...
  • Page 79 Comments Time Set slope time < > Time Set Select the level to be adjusted by Vertical To go to set up menu. See table Set Up 2-38 2-49 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 80 Select the specified line number for sync < > Lines Set PAL/SECM Standard Select Video standard NTSC Set Up To go to set up menu, see table 2-39 © 2008 RIGOL Technologies, Inc. 2-50 User’s Guide for DS1000E, DS1000D Series...
  • Page 81: Pattern Trigger (Ds1000D Series)

    If no edge is appointed, the oscilloscope will be triggered on the last edge whose code is true. 2-51 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 82 You can only appoint one code as edge. If you have appointed an edge, then appointed another edge in a different channel, and the first appointed edge will be set to X (Ignore). © 2008 RIGOL Technologies, Inc. 2-52 User’s Guide for DS1000E, DS1000D Series...
  • Page 83: Duration Trigger (Ds1000D Series)

    Acquire waveform when trigger occurred Sweep Normal When trigger occurs, acquire one waveform Single and then stop Set Up To go to set up menu, see table 2-40 2-53 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 84 X (Ignore): Don’t care. If all the channels are ignored, the oscilloscope won’t be triggered. Qualifier: A timer begins when code terms are satisfied. Duration trigger occurred in the time set by the qualifier. © 2008 RIGOL Technologies, Inc. 2-54 User’s Guide for DS1000E, DS1000D Series...
  • Page 85: Trigger Setup

    Figure 2- 63 Table 2- 39 The Trigger Set Up menu (Settings for sensitivity and holdoff) Menu Settings Comments Sensitivity Set trigger sensitivity <Sensitivity Setting> time slot before Holdoff <Holdoff Setting> another trigger event Holdoff Reset Holdoff time Reset 100ns 2-55 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 86 Table 2- 40 The Trigger Set Up menu (Settings only for holdoff) Menu Settings Comments Set time slot before another Holdoff <Holdoff Setting> trigger event Holdoff Reset Holdoff time to 100ns Reset © 2008 RIGOL Technologies, Inc. 2-56 User’s Guide for DS1000E, DS1000D Series...
  • Page 87: Trigger Holdoff

    Turn the multi function knob ( ) to change Holdoff time until waveform is stable. Pushing Trigger Hold off reset can reset the Holdoff time to its default value. 2-57 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 88: Trigger Source

    If valid triggers occur, the display becomes stable on the screen. Any factor results in the un-stability of waveforms can be detected by Auto © 2008 RIGOL Technologies, Inc. 2-58 User’s Guide for DS1000E, DS1000D Series...
  • Page 89 6div data of pre-trigger and delayed trigger can be surveyed. More data (14div) of pre-trigger and 1s delayed trigger can be surveyed by adjusting the horizontal knob. 2-59 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 90 0.1div-1.0div, which means when it sets to 1.0div, the trigger circuit will not affect any signal with peak-peak amplitude less than 1.0div, so as to avoid the influence of the noise. © 2008 RIGOL Technologies, Inc. 2-60 User’s Guide for DS1000E, DS1000D Series...
  • Page 91: To Set Up The Sampling System

    Set up memory as 512k or 1M Depth Normal Set up memory as 8k or 16k Set the insert mode to sinx/x Sinx/x Set the insert mode to line 2-61 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 92 Select Real-time acquisition to observe the single-shot or pulse signals.  Select Equ-Time to observe high frequency repetitive signals.  To reduce the displayed random noise, select Average Acquisition. And this  © 2008 RIGOL Technologies, Inc. 2-62 User’s Guide for DS1000E, DS1000D Series...
  • Page 93 Stop Acquisition: When the scope is acquiring waveforms, the waveforms is in a live status; when acquisition is stopped, frozen waveform will be displayed, the position and scale can still be adjusted by vertical control and horizontal control. 2-63 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 94 2 to 256. Peak Detect: Peak Detect mode captures the maximum and minimum values of a signal. Finds highest and lowest record points over many acquisitions. © 2008 RIGOL Technologies, Inc. 2-64 User’s Guide for DS1000E, DS1000D Series...
  • Page 95: To Set Up The Display System

    Clear from screen Infinite The sample points remain displayed until turn the persist persistence “OFF”. Turn off the persistence function Intensity Set up waveform intensity < > percentage 2-65 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 96 Adjusting waveform intensity Default setup of multi-function knob ( ) is adjusting waveform intensity. © 2008 RIGOL Technologies, Inc. 2-66 User’s Guide for DS1000E, DS1000D Series...
  • Page 97: To Store And Recall

    Go to menu for internal memory Internal operation (see table 2-48) Go to menu for external memory External operation (see table 2-49) Go to disk manage menu (see Disk Mana. table 2-50) 2-67 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 98 Go to menu for external memory External operation (see table 2-49) Go to disk manage menu Disk Mana. (see table 2-50) © 2008 RIGOL Technologies, Inc. 2-68 User’s Guide for DS1000E, DS1000D Series...
  • Page 99 Go to menu for external External memory operation(see table 2-49) Go to disk manage menu Disk Mana. (see table 2-50) 2-69 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 100 Recall waveform files and setup files Load from the internal memory location Save waveform files and setup files to Save the internal memory location Delete File Delete the selected location file (Folder) (folder) © 2008 RIGOL Technologies, Inc. 2-70 User’s Guide for DS1000E, DS1000D Series...
  • Page 101 Delete the selected location file (Folder) (folder) Recall waveform and setup from Load USB storage device. Path Back to previous folder Current directory Files File info. Figure 2- 81 File system 2-71 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 102 Figure 2- 83 Table 2- 51 The Disk Manage menu (Page 2/2) Menu Settings Comments Rename To rename a file (see table 2-52) Disk info Display disk information © 2008 RIGOL Technologies, Inc. 2-72 User’s Guide for DS1000E, DS1000D Series...
  • Page 103 Table 2- 52 The Rename menu Menu Settings Comments To move the cursor up To move the cursor down To delete chosen letter Rename the file Figure 2- 85 Rename the file 2-73 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 104 To ensure the setups being saved properly, only after the settings are changed for more than 5 seconds can the user turn off the instrument. The oscilloscope can store 10 settings permanently and can restore at anytime. © 2008 RIGOL Technologies, Inc. 2-74 User’s Guide for DS1000E, DS1000D Series...
  • Page 105: To Set Up The Utility

    (OFF) Turn off Frequency Counter Counter Turn on Frequency Counter 简体中文 繁体中文 Select languages English Language (More languages may be added in later 日本语 firmware versions) François … 2-75 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 106 Go to preference menu Self-Cal Execute Self-calibration System Info Display system information NOTE: Self-Cal: Oscilloscope will calibrate parameter of vertical system (CH1, CH2, and Ext), horizontal system and trigger system. © 2008 RIGOL Technologies, Inc. 2-76 User’s Guide for DS1000E, DS1000D Series...
  • Page 107: The I/O Setup

    Table 2- 56 The I/O Setup menu Menu Display Comments RS-232 Set RS-232 baud rate as 300, Baud 2400, 4800, 9600, 19200 or 38400. 38400 GPIB# 0-30 Set the GPIB address from 0-30. 2-77 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 108: Preference

    Sticky key: If sticky feature is turned ON, when adjusting positions (CH1, CH2, Math, Ref, Trig level and Trig offset), the object will stop at zero position until next adjustment, for the ease of getting back to initial positions. © 2008 RIGOL Technologies, Inc. 2-78 User’s Guide for DS1000E, DS1000D Series...
  • Page 109: Self-Calibration

    The self-calibration menu is displayed as Figure 2-92. Figure 2- 92 NOTE: The oscilloscope must have been working or warm-up at least 30-minutes before running self-calibration to get best accuracy. 2-79 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 110: Pass/Fail

    Output Pass Output when Pass condition detected Stop Stop test when output occur on Output Continue test when output occur Mask Setting Go to mask setting menu © 2008 RIGOL Technologies, Inc. 2-80 User’s Guide for DS1000E, DS1000D Series...
  • Page 111: Mask Setting

    Store created test mask into internal Save memory Recall mask setting file from Load internal memory Go to import/export menu (same as Imp./Exp. REF import/export menu. See table 2-10) 2-81 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 112 Go to save menu (same as REF save Save menu. See table 2-12) Load Go to load menu see table 2-63 Go to import menu. (same as REF Import import menu see table 2-14) © 2008 RIGOL Technologies, Inc. 2-82 User’s Guide for DS1000E, DS1000D Series...
  • Page 113 Table 2- 63 The Load menu Menu Settings Comments Path Switch among Path, Directory and Explorer Directory File File Load Recall the specified file NOTE: Pass/Fail function is unavailable in X-Y mode. 2-83 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 114 Before connecting to an external circuit, make sure the maximum voltage/current does not exceed 400V/100mA. The output device has no polarity limit and can be connect arbitrarily. Figure 2- 99 Schematic Diagram of Pass/Fail output © 2008 RIGOL Technologies, Inc. 2-84 User’s Guide for DS1000E, DS1000D Series...
  • Page 115: Print Setting

    Table 2- 64 The Print Set menu Menu Settings Comments Print Execute print function invert the color for print Inverted print original color Gray scale Palette Set up print color Color 2-85 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 116: Waveform Recorder

    Turn off all recorder functions Source Select record source channel P/F-OUT End Frame Set number of record frames <1-1000> Record stopped, press to (Run) Operate Start recording (Stop) Recording, press to stop © 2008 RIGOL Technologies, Inc. 2-86 User’s Guide for DS1000E, DS1000D Series...
  • Page 117 RIGOL Figure 2- 102 Table 2- 66 The Record menu (Page 2/2) Menu Settings Comments Interval Set time interval <1.00ms-1000s> 2-87 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 118 <1-1000> Current Select current frame to be played Frame <1-1000> End Frame Set End frame <1-1000> NOTE: the RUN/STOP button can also replay or continue the waveform display. © 2008 RIGOL Technologies, Inc. 2-88 User’s Guide for DS1000E, DS1000D Series...
  • Page 119 Save recorded waveform to internal Save memory location Recall recorded waveform from internal Load memory location Go to import/export menu (same as REF Imp./Exp. import/export menu. See table 2-10) 2-89 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 120 Go to save menu (same as REF save Save menu. See table 2-12) Load Go to load menu see table 2-63 Go to import menu. (same as REF Import import menu see table 2-14) © 2008 RIGOL Technologies, Inc. 2-90 User’s Guide for DS1000E, DS1000D Series...
  • Page 121: Language

    RIGOL Language The oscilloscopes have multi-language user menu, choose as your desire. Press Utility→Language to select the language. Figure 2- 108 The Language menu 2-91 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 122: To Measure Automatically

    Select to measure voltage parameter Time Select to measure time parameter Clear Clear measurement result on screen Display Turn off all measurement result Turn on al measurement result © 2008 RIGOL Technologies, Inc. 2-92 User’s Guide for DS1000E, DS1000D Series...
  • Page 123: Voltage Measurements

    Table 2- 75 The Voltage Measurements menu (Page 3/3) Menu Settings Comments Measure overshoot in percentage Overshoot of an edge Measure preshoot in percentage of Preshoot an edge 2-93 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 124: Time Measurements

    Maximum 3 results could be displayed at the same time. When there is no room, the next new measurement result will make the previous results moving left out of screen. © 2008 RIGOL Technologies, Inc. 2-94 User’s Guide for DS1000E, DS1000D Series...
  • Page 125 Clear the measure values: press Clear and all of the auto measure values would disappear from the screen. Clear measured data from screen Turn on/off the mass measurement Figure 2- 117 Measurement display 2-95 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 126: The Automatic Measurement Of Voltage Parameters

    Preshoot: Defined as (Vmin-Vbase)/Vamp, useful for square and pulse waveforms. Average: The arithmetic mean over the entire waveform. Vrms: The true Root Mean Square voltage over the entire waveform. © 2008 RIGOL Technologies, Inc. 2-96 User’s Guide for DS1000E, DS1000D Series...
  • Page 127: The Automatic Measurement Of Time Parameters

    : The delay between the two channels at the rising edge. Delay 1→2 : The delay between the two channels at the falling edge. +Duty: +Duty Cycle, defined as +Width/Period -Duty: -Duty Cycle, defined as -Width/Period 2-97 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 128: To Measure With Cursors

    These cursors demonstrate the electrical meanings of these measurements. NOTE: The Auto Measure mode for Cursor measuring will take no effect without automatic measurements. © 2008 RIGOL Technologies, Inc. 2-98 User’s Guide for DS1000E, DS1000D Series...
  • Page 129: Manual Mode

    3. Select the cursors type by pressing soft button as Cursor→Type→X or Y. 4. Move the cursors to adjust the increment between the cursors: (Details in the following Table) 2-99 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 130 Cursor 2 value: Binary. NOTE: The values will be automatically displayed on the right upper corner of screen when the cursor function menu is hidden or displaying other menus. © 2008 RIGOL Technologies, Inc. 2-100 User’s Guide for DS1000E, DS1000D Series...
  • Page 131 Cursor X: Cursor X appears as vertical lines on the display to measure horizontal parameters. Usually it indicates the time of trigger excursion. When the source is set as FFT, X means frequency. 2-101 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 132: Track Mode

    Cursor Operation Cursor A Turn the multifunctional knob ( ) to move cursor A horizontally Cursor B Turn the multifunctional knob ( ) to move Cursor B horizontally © 2008 RIGOL Technologies, Inc. 2-102 User’s Guide for DS1000E, DS1000D Series...
  • Page 133 △ units in seconds. X), units in Hz, kHz, MHz, GHz. △ Vertical space between cursor 1 and 2 ( Y): Voltage between cursors, units in V. △ 2-103 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 134: Auto Mode

    Auto Measure Mode of Cursor Measurement There will be no cursor display if no parameters are chosen in Measure menu. The oscilloscope could move cursor automatically to measure 20 parameters in Measure menu. © 2008 RIGOL Technologies, Inc. 2-104 User’s Guide for DS1000E, DS1000D Series...
  • Page 135: To Use Run Control Buttons

    Press to display the waveform’s falling edge and measure its fall Fall Edge time automatically Press to cancel all the Auto Set actions, the oscilloscope will (Cancel) recover to its previous status 2-105 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 136 In STOP status, the volts/div and horizontal time base can be adjusted in a fixed limit. That is, to zoom in/out the signal in vertical and horizontal directions. © 2008 RIGOL Technologies, Inc. 2-106 User’s Guide for DS1000E, DS1000D Series...
  • Page 137: Chapter 3 : Application & Examples

    Press Time→Freq to select frequency measurements and the result will be displayed on the screen. NOTE: The frequency, period, and peak-to-peak measurements are shown on the screen and are updated periodically. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 138: Example 2: View A Signal Delay Caused By A Circuit

    (3) Press Delay 1→2 to display the result on the screen. You can see the change of the waveform in the following figure: Figure 3- 1 Delay of the signals © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 139: Example 3: Capture A Single-Shot Signal

    It is useful to observe the waveform before the occurrence of the noise. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 140: Example 4: To Reduce The Random Noise On A Signal

    8 kHz. Use LF Reject to remove low frequency signals such as power line noise from the trigger path. 4. To reduce the noise by setting the acquisition type and adjust the waveform intensity. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 141 (2) To reduce the noise it can also be achieved by reducing the intensity of the display. NOTE: It is normal that the refresh rate will slow down when the average acquisition mode is ON. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 142: Example 5: Making Cursor Measurements

    ) knob to place cursor B on the second peak of the wave. Observe the delta in time and frequency displayed on the screen. Figure 3- 4 Waveform display © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 143 Observe the following measurements in the cursor menu: (See Figure 3-5) The delta voltage (peak-to-peak voltage of the waveform)  The voltage at Cursor 1  The voltage at Cursor 2  Figure 3- 5 Waveform display © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 144: Example 6: The Application Of The X-Y Operation

    8. Adjust the vertical knobs to a desirable waveform display. 9. Apply the Ellipse method to observe the phase difference between the two channels. (See Figure 3-6) © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 145 If the main axis of the ellipse is at I and III quadrant,  must be in the range of (0~π/2) or (3π/2~2π). If the main axis is at II and IV quadrant,  must be in the range of (π/2~π) or (π~3π/2). © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 146: Example 7: Triggering On A Video Signal

    The oscilloscope triggers on the Odd field or Even field. To avoid confusion when Odd field and Even field trigger simultaneously, choose Odd field or Even field as in step 5 above. © 2008 RIGOL Technologies, Inc. 3-10 User’s Guide for DS1000E, DS1000D Series...
  • Page 147 8. Turn the horizontal knob to observe a complete waveform on the screen. Figure 3- 8 Waveform display 3-11 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 148: Example 8: Fft Cursor Measurement

    3. Press Source to select FFT. 4. Turn ( ) knob to move the cursor to an interesting point. Figure 3- 9 The example of Cursor measuring FFT amplitude © 2008 RIGOL Technologies, Inc. 3-12 User’s Guide for DS1000E, DS1000D Series...
  • Page 149 RIGOL Figure 3- 10 The example of Cursor measuring FFT frequency 3-13 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 150: Example 9: Pass/Fail Test

    Create Mask to create a new mask. Press Output to select the expected outputting waveforms. Press Operate to start the test. Figure 3- 11 Waveform display © 2008 RIGOL Technologies, Inc. 3-14 User’s Guide for DS1000E, DS1000D Series...
  • Page 151: Example 10: Triggering On A Digital Signal

    5. Press Sweep to select trigger mode to: Auto, Normal, or Single 6. Press Set Up to adjust the Holdoff time. Figure 3- 12 Pattern Trigger on a digital signal 3-15 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 152 7. Press Sweep to select trigger mode to: Auto, Normal, or Single 8. Press Set Up to adjust the Holdoff time. Figure 3- 13 Duration Trigger on a digital signal © 2008 RIGOL Technologies, Inc. 3-16 User’s Guide for DS1000E, DS1000D Series...
  • Page 153 RIGOL Figure 3- 14 Duration Trigger on a digital signal 3-17 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 155: Chapter 4 : Prompt Messages & Troubleshooting

    Time/div at limit: The horizontal scale is up to the highest horizontal resolution under the Equivalent-sampling mode. Memory position at limit: The Memo Offset is at the full range of the memory. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 156 Failed upgrade: Failed to upgrade from a USB drive Files are covered: The original file will be replaced by a new file when store a new file to the memory. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 157: Troubleshooting

    (2) Ensure the power switch is turned on. (3) After the above inspection, restart the oscilloscope. (4) If the problem still remains, please contact RIGOL for help. 2. After the signal acquisition the waveform does not appear: (1) Check the probes connected with the signals.
  • Page 158 (1) The time base setting maybe is too slow. Turn the horizontal knob to increase horizontal resolution to improve the display. (2) Maybe the display Type is set to “Vectors” set it to “Dots” mode to improve the display. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 159: Chapter 5 : Specifications

    Do perform the Self Cal operation, accessible through the Utility menu, if the  operating temperature changes by more than 5°C. All specifications are guaranteed unless noted “typical”.  © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 160: Specifications

    Double channel 500MSa/s 8 Kpts 512Kpts Or lower Scan speed Range 2ns/div~50s/div, DS1102X (Sec/div) 5ns/div~50s/div, DS1052X 1-2-5 Sequence Sample Rate and ± 50ppm (over any 1ms time interval) Delay Time © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 161 Average of ≥16 Waveforms with vertical position not at zero: ± [DC Gain Accuracy× (reading+vertical position)+(1% of vertical position) + 0.2div] Add 2mV for settings from 2mV/div to 200 mV/div © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 162: Alternate Trigger

    (>, <, =) Positive slope, (>, <, =) negative slope Trigger condition Time setting 20ns~10s Alternate Trigger Trigger on CH1 Edge, Pulse, Video, Slope Trigger on CH2 Edge, Pulse, Video, Slope Pattern Trigger © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 163 Preshoot, Freq, Period, Rise Time, Fall Time, +Width, -Width, +Duty, -Duty, Delay1→2 , Delay1→2 [1] For DS1000D Series; [2] When sampling is 1GSa/s, only single channel can be used. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 164: General Specifications

    +35℃~ +40℃: ≤60% relative humidity Altitude Operating 3,000 m or below Non-operating 15,000 m or below Mechanical Size Width 303mm Height 154mm Depth 133 mm Heavy Without package 2.4 kg Packaged 3.8 kg © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 165: Calibration Interval

    RIGOL IP Degree IP2X Calibration Interval The recommended calibration interval is one year © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 167: Chapter 6 : Appendix

    A Warranty Card  Optional Accessories: BNC Cable  RS232 Cable  DS1000E, DS1000D soft carrying case  All accessories (standard and optional) are available by contacting your local RIGOL office. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 168: Appendix B: Warranty

    To get repair service or obtain a copy of the whole warranty statement, please contact with your nearest RIGOL sales and service office. RIGOL do not provide any other warranty items except the one being provided by this summary and the warranty statement. The warranty items include but not being subjected to the hint guarantee items related to tradable characteristic and any particular purpose.
  • Page 169: Appendix C: Care And Cleaning

    Use a soft cloth dampened with water to clean the instrument. NOTE: To avoid damage to the surface of the instrument or probes, do not use any abrasive or chemical cleaning agents. © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 170: Appendix D: Contact Rigol

    RIGOL Technologies, Inc. 156# CaiHe Village, ShaHe Town, ChangPing District, Beijing, China Post Code: 102206 Overseas: Contact the local RIGOL distributors or sales office. For the latest product information and service, please visit our website: www.rigolna.com © 2008 RIGOL Technologies, Inc.
  • Page 171: Index

    Front panel ....... 1-2 Save and Recal ......2-67 Functional check ......1-6 Self- Calibration ......2-79 GND Coupling ......2-5 Single Trigger ......2-59 Hamming Window..... 2-15 Slope Trigger......2-45 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...
  • Page 172 Vertical Window ....... 1-12 Trigger Control......1-16 Video Signal ......3-10 Trigger System ......2-36 Video Trigger ......2-41 Troubleshooting ......4-3 X-Y ......... 2-32 UTILITY ........2-75 Y-T ......... 2-32 © 2008 RIGOL Technologies, Inc. User’s Guide for DS1000E, DS1000D Series...

This manual is also suitable for:

Ds1000e seriesDs1102eDs1052eDs1102dDs1052d