Measurement Setting - Rigol DS6062-DEMO User Manual

Digital oscilloscope
Hide thumbs Also See for DS6062-DEMO:
Table of Contents

Advertisement

Measurement Setting

1. Source Selection
Press MEAS  Source to select the desired channel for measurement
(CH1-CH4). The color of the parameter icon under the MENU at the left of the
screen will change with the source selected.
2. Measurement Range
Press MEAS  Meas.Range  Meas.Range to select "Screen Region" or
"Cursor Region" for measurement. When "Cursor Region" is selected, two cursor
lines appear on the screen. At this point, press Cursor A and Cursor B and use
to adjust the position of the two cursor lines respectively so as to determine
the measurement range.
3. Delayed Measurement Setting
Specify the source A and source B in the measurement items "Delay A→B "
and "Delay A→B ".
Press MEAS  Meas Setting  Type  "Delay" and then press Source A
and Source B to set the two channel sources (CH1-CH4) of the delayed
measurement respectively.
4. Phase Measurement Setting
Specify the source A and source B in the measurement items "Phase A→B "
and "Phase A→B ".
DS6000 User's Guide