To Invert A Waveform; Figure 5.2 Switched Input Impedance - Rigol HDO4000 Series User Manual

Digital oscilloscope
Hide thumbs Also See for HDO4000 Series:
Table of Contents

Advertisement

1 MΩ: In this mode, the input impedance of the oscilloscope is very high, and
the current flowed from the circuit under test can be ignored.
50 Ω: It makes the oscilloscope match with the device whose input impedance is
50 Ω.
After the oscilloscope switches to the 50 Ω mode, the circuit diagram in the Vertical
menu will also be changed, as shown in the figure below.
TIP
After the oscilloscope recognizes the probe automatically, the input impedance will also
be auto recognized to save you from configuring it manually.
Setting the input impedance affects the ranges of vertical scale and offset for the
specified channel.
5.7

To Invert a Waveform

Click or tap the channel status label at the bottom of the screen, and then the
Vertical menu is displayed. Then click or tap the Invert on/off switch to enable or
disable the waveform invert function.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.

Figure 5.2 Switched Input Impedance

Vertical System
HDO4000 User Guide