Summary of Contents for Rigol DHO1000 Series

  • Page 1 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 2 • RIGOL products are covered by P.R.C. and foreign patents, issued and pending. • RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at the company's sole decision. • Information in this publication replaces all previously released materials.
  • Page 3: Table Of Contents

    Product Overview......................4.4.1 Front Panel Overview....................4.4.2 Rear Panel Overview....................24 4.4.3 User Interface Overview..................26 Touch Screen Gestures.....................28 4.5.1 Tap...........................28 4.5.2 Pinch&Stretch......................4.5.3 Drag..........................29 Parameter Setting Method..................... 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 4 Horizontal Expansion Reference...................61 Roll Mode..........................Mode..........................62 Triggering the Oscilloscope....................Trigger Source........................65 Trigger Level........................Trigger Mode........................67 Trigger Coupling.........................68 Trigger Holdoff........................69 Noise Rejection........................69 Trigger Type......................... 8.7.1 Edge Trigger........................ 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 5 Delay and Phase Parameters..............10.2.1.4 Voltage Parameters..................10.2.1.5 Other Parameters....................145 10.2.2 Select the Measurement Item................145 10.2.3 Measurement Settings..................10.2.4 Remove the Measurement Results..............10.3 Cursor Measurements....................151 10.3.1 Manual Mode......................1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 6 14.2.1 Source Setting......................14.2.2 To Set Data Package....................181 14.2.3 Display-related Settings..................182 14.2.4 Event Table.........................182 14.3 I2C Decoding........................14.3.1 Source Setting......................14.3.2 Display-related Settings..................185 14.3.3 Event Table.........................185 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 7 Load.........................214 19.1 To Enter the Storage Menu..................19.2 To Save a File........................214 19.2.1 Save Image........................ 19.2.2 Save Wave........................216 19.2.3 Save Setup......................... 19.2.4 Binary Data Format (.bin)..................1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 8 Remote Control via LAN....................21.3 Remote Control via GPIB....................Troubleshooting........................Appendix............................244 23.1 Appendix A: Options and Accessories..............23.2 Appendix B: Warranty....................23.3 Appendix C: Factory Settings..................1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 9 Figure 5.3 Probe Setting Menu ...................46 Figure 5.4 Zero Offset ......................Figure 6.1 Horizontal Menu ....................Figure 6.2 Zoom Mode ......................Figure 7.1 Horizontal Menu ....................1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 10 Figure 8.22 Nth Edge Trigger ....................97 Figure 8.23 Nth Edge Trigger Setting Menu ..............Figure 8.24 Schematic Diagram of RS232 Protocol ............. Figure 8.25 RS232 Trigger Setting Menu ..............1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 11 Figure 9.13 Operation Result Display Window ............Figure 10.1 Time Parameters .....................139 Figure 10.2 Delay and Phase Parameters ..............142 Figure 10.3 Voltage Parameters ..................Figure 10.4 Vertical Measurement Items ..............1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 12 Figure 14.10 SPI Serial Bus ....................187 Figure 14.11 SPI Decoding Menu ..................Figure 14.12 SPI Decoding Event Table ................. Figure 14.13 CAN Decoding Menu ................. Figure 14.14 Sample Position ....................192 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 13 Figure 20.3 Key Test Interface ................... Figure 20.4 Touch Screen Test Interface ................237 Figure 21.1 Search for the Available Device ..............241 Figure 21.2 Confirm the Available Device ..............241 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 14 Table 19.1 BIN File Format ....................Table 19.2 File Header ......................220 Table 19.3 Waveform Header ....................220 Table 19.4 Waveform Data Header ................. Table 23.2 Factory Settings ....................1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 15: Safety Requirement

    Do not insert objects into the air outlet, as doing so may cause damage to the instrument. • Use the Proper Fuse. Please use the specified fuses. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 16 • Do Not Operate With Suspected Failures. If you suspect damage occurs to the instrument, have it inspected by RIGOL authorized personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.
  • Page 17: Safety Notices And Symbols

    Measurement Category Measurement Category This instrument can make measurements in Measurement Category I. WARNING This instrument can only be used for measurements within its specified measurement categories. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 18: Ventilation Requirement

    Working Environment Temperature Operating: 0℃ to +50℃ Non-operating: -30℃ to +60℃ Humidity • Operating: Below +30℃: ≤90% RH (without condensation) +30℃ to +40℃: ≤75% RH (without condensation) 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 19 The pollution has no effect. For example, a clean room or air-conditioned office environment. • Pollution Degree 2: Normally only nonconductive pollution occurs. Temporary conductivity caused by condensation is to be expected. For example, indoor environment. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 20: Care And Cleaning

    Please contact your local authorities for disposal or recycling information. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 21: Product Features

    • Standard USB Device & Host, LAN, and HDMI interfaces DHO1000 series digital oscilloscope is designed to meet the designing, debugging, and testing requirements of the mainstream oscilloscope market. Adopting the brand-new chipset "Centaurus" developed by RIGOL, this series achieves a fast...
  • Page 22: Document Overview

    Document Overview Document Overview This manual gives you a quick overview of the front and rear panel, user interface as well as basic operation methods of DHO1000 series. Publication Number UGA32101-1110 Software Version Software upgrade might change or add product features. Please acquire the latest...
  • Page 23 Vertical SCALE Knob Content Conventions in this Manual DHO1000 series digital oscilloscope includes the following models. Unless otherwise specified, this manual takes the 4-channel model DHO1104 as an example to illustrate the basic operation methods of DHO1000 series. Model Max. Analog Bandwidth...
  • Page 24: Quick Start

    The consigner or carrier shall be liable for the damage to the instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the instrument. 2. Inspect the instrument...
  • Page 25: To Prepare For Use

    Flip in the tabs for storage or transporting when not using the instrument. Please see the figure below. Figure 4.3 Flipping in/out Tabs 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 26: Connecting To Power

    "Are you sure to shutdown?" is displayed. Click or tap OK to shut down the instrument. Press twice to directly shut down the instrument. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 27: Replacing The Fuse

    To avoid electric shock, please make sure that the instrument is powered off and disconnected from the power before replacing the fuse. Also, please make sure the fuse is consistent with the required fuse rating. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 28: Setting The System Language

    Setup > Language to select the system language. 4.3.6 Connecting Probes RIGOL provides passive probes for DHO1000 series. For specific probe models, please DHO1000 series Data Sheet . For detailed technical information of the probes, refer to please refer to the corresponding Probe User Guide.
  • Page 29: Figure 4.7 Using The Compensation Signal

    To avoid electric shock when using the probe, please make sure that the insulated wire of the probe is in good condition. Do not touch the metallic part of the probe when the probe is connected to high voltage source. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 30: Probe Compensation

    "Perfectly compensated" waveform shown in the figure above. Product Overview The following content provides a description of the front and rear panels as well as the user interface of the DHO1000 series by taking 4-channel model DHO1104 as an example. 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 31: Front Panel Overview

    - Non-menu operation: When not operating on the menu, you can rotate the knob 1 to adjust the waveform brightness. When a cursor, decoding, Math waveform, or reference 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 32 6. Touch Lock Key Press once to disable the touch screen function; press it again to enable the touch screen function. 7. Common Tools Keys 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 33 9. Trigger Controls is the trigger setup key. Press this key to access the Trigger menu. For Triggering the Oscilloscope details, please refer to 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 34 Math key. Press this key to access the Math menu. Math operations include A+B, A-B, A× B, A/B, FFT, etc. You can also set the Math label. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 35 Press this key to access the Horizontal menu. You can set the oscilloscope's time base mode, acquisition mode, memory depth, etc. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 36 USB-GPIB module. - USB storage device: imports or exports data (software update, waveform, setup, or captured image). - USB-GPIB module: extends the GPIB interface for RIGOL instruments that integrates the USB HOST interface but not the GPIB interface. 15. Power Switch Press the switch to power on/off the oscilloscope.
  • Page 37: Rear Panel Overview

    4. 10 MHz REF OUT BNC connector that can output the 10 MHz clock signal generated by the internal crystal oscillator inside the instrument. 5. HDMI 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 38 If you need to replace the fuse, use only the specified fuse. 10. Security Lock Hole Use a standard PC/laptop lock cable to secure the oscilloscope to a work bench or other location. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 39: User Interface Overview

    Displays the current sample rate and memory depth. Click or tap this label to enter the horizontal setting menu. 5. Horizontal Delay Label Displays the current horizontal position. Click or tap this label to enter the horizontal setting menu. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 40 11. Math Labels Displays the on/off state, operation type, and vertical scale. 12. Channel Labels - Display the on/off status of CH1~CH4. - Display the channel coupling mode. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 41: Touch Screen Gestures

    Tap the close button at the upper-right corner of the message box to close the prompt window. • Tap other windows on the touch screen and operate on the windows. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 42: Pinch&Stretch

    Use one finger to select the object, and then drag the object to a destination place, as shown in the figure below. With the drag gesture, you can perform the following operation: • Drag the waveform to change its position or scale. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 43: Parameter Setting Method

    When naming a label, this instrument supports Chinese/English input method. The following part introduces how to input Chinese and English characters with the Chinese/English input method. • Input English Characters 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 44: Figure 4.16 English Input Interface

    If the letter keypad is displayed, you need to click or tap the numeric switchover key to switch to the numeric keypad, and input numbers or 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 45: Figure 4.17 Chinese Input Interface

    First check the input method type. If it shows "中/En", then go to Step 2; if it shows "En/中", click or tap the input method switchover key to switch to "中/En" (Chinese input method). 2. Clear the name input area 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 46 After completing the input operation, click or tap "OK". Input a String When naming a file or folder, you need to input a string with the keypad. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 47: Figure 4.18 String Keypad

    During the name input process, you can modify or delete the unwanted character if necessary. To delete the characters that have been entered, click or tap the 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 48: Figure 4.19 Numeric Keypad

    • Delete the parameter value that has been input; • Set the parameter value to a maximum or minimum value; 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 49: Using The Security Lock

    The built-in help system provides instructions for all the front-panel keys and their corresponding menu keys. Click or tap > Help to access the "Help" menu. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 50: Viewing The Option Information And Install The Option

    1. Obtain an option license a. Log in to the RIGOL official website click SERVICE CENTRE > License Activation to enter the "Registered product license code" interface. b. In the software license registration interface, input the correct key, serial number (click or tap >...
  • Page 51: Vertical System

    CH1 on or off. When CH1 is activated, its status label at the bottom of the screen is as shown in the figure below. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 52 In addition, you can slide down the channel label to disable the channel. If CH1 is disabled, its status label is as shown in the figure below. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 53: To Adjust The Vertical Scale

    OFF (coarse adjustment). The default setting is OFF. You can also press the Vertical SCALE knob to toggle between "coarse adjustment" and "fine adjustment". 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 54: To Adjust The Vertical Offset

    You can also click or tap the input field to input a specific value with the displayed numeric keypad. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 55: To Specify Channel Coupling

    This series oscilloscope supports the bandwidth limit function. Setting the bandwidth limit can reduce the noises in the displayed waveforms. For example, the signal under test is a pulse with high frequency oscillation. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 56: To Specify Input Impedance

    Vertical menu is displayed. Then click or tap the Invert on/off switch to enable or disable the waveform invert function. When "ON" is selected, the channel label is as shown in the figure below. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 57: To Set Probe

    Click or tap the channel status label at the bottom of the screen. Then the menu Vertical is displayed. Then click or tap Probe to enter the Probe setting menu, as shown in the figure below. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 58: Figure 5.3 Probe Setting Menu

    (display amplitude of the signal: actual amplitude of the signal) 0.001X 0.001:1 0.002X 0.002:1 0.005X 0.005:1 0.01X 0.01:1 0.02X 0.02:1 0.05X 0.05:1 0.1X 0.1:1 0.2X 0.2:1 0.5X 0.5:1 1X (default) 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 59 0.00 s. Go Back to the Vertical Menu In the Probe setting menu, click or tap Vertical to go back to the Vertical menu. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 60: To Specify The Amplitude Unit

    (zero offset). This series oscilloscope allows you to set a delay time for calibrating the zero offset of the corresponding 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 61: To Turn The Channel Label Display On/Off

    You can also click or tap the label input field to input a specific name for the channel label with the pop-up keypad. For how to use the keypad, please refer to descriptions in Parameter Setting Method 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 62: Horizontal System

    2.00 ns/div to 1.00 ks/div. While you change the horizontal time base, the displayed waveforms of all channels are expanded or compressed horizontally relative to the selected time reference 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 63: To Adjust The Horizontal Position

    (negative) value. Changing the horizontal position moves the waveform trigger points and the displayed waveforms of all channels horizontally. The horizontal position displayed in 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 64: Zoom Mode (Delayed Sweep)

    Zoomed Scale: Rotate the corresponding multipurpose knob or use the icons at the right of the Scale input field to increase or decrease the time base for the 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 65: Figure 6.2 Zoom Mode

    Compared with the main time base, the zoomed time base has higher horizontal resolution. The zoomed time base should be smaller than or equal to the main time base. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 66: Acquisition System

    On the other hand, the higher the number of averages, the slower the response of the displayed waveform to waveform changes. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 67 Click or tap the Display mode drop-down button to select the display mode of the acquired data. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 68: Figure 7.2 Display Modes

    In UltraAcquire mode, the functions mentioned above (except for roll mode and slow sweep mode) are disabled. When roll mode or slow sweep mode is enabled, the acquisition mode is automatically switched to "Normal". 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 69: Sampling Mode

    All channels mode: For two-channels models, if all two channels are enabled, it is called all channels mode; for four-channel models, if any three channels or all four channels are enabled, it is called all channels mode. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 70: Memory Depth

    Memory depth refers to the number of points that the oscilloscope can store in a single trigger acquisition. It reflects the storage capability of the acquisition memory. DHO1000 series oscilloscope provides up to 50 Mpts memory depth (100 Mpts optional).
  • Page 71: Figure 7.3 Memory Depth

    When two of the channels are enabled, the memory depths available include Auto, 1 kpts, 10 kpts, 100 kpts, 1 Mpts, 10 Mpts, 25 Mpts, and 50 Mpts (optional). 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 72: Horizontal Expansion Reference

    The roll mode causes the waveform to move across the screen from right to left. It allows you to view the acquired data without waiting for a complete acquisition. In Horizontal menu, select "Auto" or "OFF" in Roll. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 73: Xy Mode

    In the "Horizontal" menu, tick XY to enable the XY mode. Configure the XY Mode Click or tap at the upper-right corner of the XY display window to enter the XY configuration menu. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 74: Figure 7.4 Xy Menu

    To Set the Screen Grid • Grid: Please refer to NOTE Advanced settings are not available for now. The current settings can produce the best display. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 75: Figure 7.5 Measurement Schematic Diagram Of Phase Deviation

    The XY mode can be used to measure the phase deviation occurred when the signal under test passes through a circuit network. Connect the oscilloscope to the circuit to monitor the input and output signals of the circuit. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 76: Triggering The Oscilloscope

    AC power frequency. For example, it can stably trigger the waveform output from the transformer of a transformer substation. It is mainly used in related measurements of the power industry. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 77: Trigger Level

    Runt trigger, Window trigger), you need to click or tap 90% for Level A and 10% for Level B to make the level move within the range of the waveform amplitude. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 78: Trigger Mode

    This trigger mode can be used when the signal levels are unknown, when the DC signals should be displayed, or when trigger conditions occur often enough that forced triggers are unnecessary. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 79: Trigger Coupling

    When you adjust the trigger level, you can only see the changes of the trigger level values in the trigger information label at the top of the screen. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 80: Trigger Holdoff

    8 ns (default) to 10 s. Noise Rejection Noise rejection can reject the high frequency noise in the signal and reduce the possibility of triggering the oscilloscope by mistake. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 81: Trigger Type

    The information will change based on the trigger settings. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 82 For details, refer to descriptions in Level . The current trigger level is displayed in the trigger information label at the top of the screen. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 83: Pulse Width Trigger

    Figure 8.5 Pulse Width Trigger Setting Menu After a trigger type is selected, the current trigger setting information (including trigger type, trigger source, and trigger level) is displayed in the trigger information 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 84 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 85: Slope Trigger

    A and B with the falling edge. See the figure below. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 86: Figure 8.6 Positive Slope Time/Negative Slope Time

    Click or tap Source drop-down button to select CH1-CH4. For details, refer to Source . The selected trigger source is indicated in the trigger information label at the top of the screen. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 87 Upper and Lower respectively to set the upper limit value and the lower limit value with the pop-up numeric keypad. You can also use the corresponding 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 88: Video Trigger

    NTSC (National Television Standards Committee), PAL (Phase Alternating Line), or SECAM (Sequential Couleur A Memoire). Trigger Type Click or tap the Type drop-down button to select "Video". 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 89: Figure 8.8 Video Trigger Setting Menu

    In the Polarity item, select the desired polarity: positive polarity ( ) or negative polarity ( Video Standard Click or tap the drop-down button of Standard to select the desired video standard. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 90: Table 8.1 Video Standard

    "NTSC" or "PAL/SECAM". Trigger Mode In Sweep, select Auto, Normal, or Single as the trigger mode. For details, refer to Trigger Mode descriptions in 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 91: Pattern Trigger

    In the trigger debugging process of video signals, the frequency in different part of the signal can be reflected by a different brightness, as RIGOL's digital oscilloscope provides the intensity graded color display function. Experienced users can quickly judge the signal quality and discover abnormalities during the debugging process.
  • Page 92: Figure 8.10 Pattern Trigger Setting Menu

    X: sets the pattern of the channel selected to "X", i.e. this channel is not used as a part of the pattern. When all channels in the pattern are set to "X", the oscilloscope will not trigger. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 93: Duration Trigger

    This pattern is a logical "AND" combination of the channels. Each channel can be set to 1 (high), 0 (low), or X (don't care). The 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 94: Figure 8.11 Duration Trigger

    Figure 8.12 Duration Trigger Setting Menu After a trigger type is selected, the current trigger setting information (including trigger type, trigger source, and trigger level) is displayed in the trigger information 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 95 1.01 ns to 10 s. Click or tap the input field of Lower to set the lower limit of the duration of the pattern, and the range is from 1 ns to 9.9 s. You 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 96: Timeout Trigger

    Figure 8.13 Time Out < T Time Out Figure 8.13 Timeout Trigger 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 97: Figure 8.14 Timeout Trigger Menu

    Rising: starts timing when the rising edge of the input signal passes through the trigger level. • Falling: starts timing when the falling edge of the input signal passes through the trigger level. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 98: Runt Trigger

    Lower Limit of the Trigger Level Negative Runt Pulse Figure 8.15 Runt Trigger Trigger Type Click or tap the drop-down button of Type to select "Runt". 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 99: Figure 8.16 Runt Trigger Setting Menu

    : triggers on the negative runt pulse. Trigger Condition Set the Runt trigger condition in the When item. • None: indicates not setting the trigger condition of Runt trigger. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 100 Press the trigger level knob to switch between three modes: "only adjust Level A", "only adjust Level B", and "adjust Level A and Level B at the same time". 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 101: Window Trigger

    Only when you select the channel that has signal inputs as the trigger source, can you obtain a stable trigger. Edge Type Select the input signal edge (in the Slope item) on which the oscilloscope triggers. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 102 Press the trigger level knob to switch between three modes: "only adjust Level A", "only adjust Level B", and "adjust Level A and Level B at the same time". 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 103: Delay Trigger

    The information will change based on the trigger settings. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 104 < >: triggers when the time difference (∆T) between the specified edges of Source A and Source B is greater than the lower limit of the preset time and 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 105: Setup/Hold Trigger

    The oscilloscope triggers when the setup time or hold time is smaller than the preset time. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 106: Figure 8.20 Setup/Hold Trigger

    The information will change based on the trigger settings. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 107 You can also use the corresponding multipurpose knob to set the values. Trigger Mode In Sweep, select Auto, Normal, or Single as the trigger mode. For details, refer to Trigger Mode descriptions in 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 108: Nth Edge Trigger

    Edge Type= Edge Num=2 P<Idle Time<M Figure 8.22 Nth Edge Trigger Trigger Type Click or tap the drop-down button of Type to select "Nth Edge". 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 109: Figure 8.23 Nth Edge Trigger Setting Menu

    Rising: triggers on the rising edge of the input signal when the voltage level meets the specified trigger level. • Falling: triggers on the falling edge of the input signal when the voltage level meets the specified trigger level. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 110: Rs232 Trigger

    RS232 signal. 1 bit 5~8 bits 1 bit 1~2 bits Start Bit Data Bit Check Bit Stop Bit Figure 8.24 Schematic Diagram of RS232 Protocol 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 111: Figure 8.25 Rs232 Trigger Setting Menu

    Select the polarity of data transmission in the Polarity item. It can be set to "Positive" or "Negative" Trigger Condition Set the desired trigger condition in the When item. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 112 In Sweep, select Auto, Normal, or Single as the trigger mode. For details, refer to Trigger Mode descriptions in Trigger Parameter Setting You can refer to Noise Rejection to set the noise rejection under this trigger type. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 113: I2C Trigger

    Figure 8.26 Sequence Diagram of I2C Protocol Trigger Type Click or tap the drop-down button of Type to select "I2C". 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 114: Figure 8.27 I2C Trigger Setting Menu

    Stop: triggers when SDA data transitions from low level to high level while SCL is high level. • Restart: triggers when another start condition occurs before a stop condition. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 115: Figure 8.28 Binary Format Setting

    Click or tap the input field of Data, and then the "Format" interface is will displayed. You can select "Bin" or "Hex" data format. Figure 8.28 Binary Format Setting 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 116: Figure 8.29 Hexadecimal Format Setting

    Click or tap the input field of Level B to input the level of SDA with the pop-up numeric keypad. You can also use the corresponding multipurpose knob to set 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 117: Spi Trigger

    The information will change based on the trigger settings. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 118 Click or tap the input field of DataBits, and then use the pop-up numeric keypad to set the number of bits in the serial data string. You can also use the corresponding 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 119: Can Trigger

    (e.g. Answer Error, Check Error, Format Error, etc.) of the CAN signal. The data frame format of the CAN bus is as shown in the figure below. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 120: Figure 8.32 Data Frame Format Of The Can Bus

    Click or tap the drop-down button of Source to select CH1-CH4. For details, refer to Trigger Source descriptions in The selected trigger source is indicated in the trigger information label at the top of the screen. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 121: Figure 8.34 Sample Position

    Click or tap the drop-down button of When to select the desired trigger condition. • SOF: triggers at the start of a frame. • EOF: triggers at the end of a frame. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 122 In Sweep, select Auto, Normal, or Single as the trigger mode. For details, refer to descriptions in Trigger Mode Trigger Parameter Setting You can refer to Noise Rejection to set the noise rejection under this trigger type. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 123: Lin Trigger

    The information will change based on the trigger settings. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 124: Figure 8.37 Sample Position

    Click or tap the input field of Sample Position to set it by using the pop-up numeric keypad. You can also use the corresponding multipurpose knob to set the value. The settable range is from 10% to 90%. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 125 For details, refer to descriptions in Level . The current trigger level is displayed in the trigger information label at the top of the screen. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 126: Trigger Output Connector

    If "PassFail" is selected in Aux Out, the instrument can output a pulse from the [AUX OUT] connector when a pass/failed event is detected during the pass/fail test. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 127: Math Operation

    This oscilloscope provides four math operations: Math1, Math2, Math3, and Math4. In the Math menu, you can select the desired math operation type by clicking or 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 128: Arithmetic Operation

    A÷B divides the waveform voltage values of signal source A by that of source B point by point and displays the results. It can be used to analyze the Multiple relation of the two channels waveforms. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 129: Figure 9.3 Arithmetic Operation Menu

    The source and the vertical scale parameters are displayed at the top of the window, as shown in the figure below. Figure 9.4 Operation Result Display Window 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 130 Invert is used to enable or disable the inverted display of the waveform. For setting To Invert a Waveform methods, refer to descriptions in Waveform This oscilloscope provides Main and Zoom for the measurement range. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 131: Function Operation

    Sqrt: calculates the square roots of the selected source point by point and displays the results. • Lg (Base 10 Exponential): calculates the base 10 exponential of the selected source point by point and displays the results. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 132: Figure 9.5 Function Operation Menu

    Click or tap the Operation on/off switch to enable or disable the display of the operation result window. The source and the vertical scale parameters are displayed at the top of the window, as shown in the figure below. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 133: Figure 9.6 Operation Result Display Window

    • Close the menu and then adjust the vertical scale with the pinch&stretch gesture on the touch screen. You can also rotate the corresponding multipurpose knob 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 134 • Center: When the vertical scale is changed, the math operation waveform will expand or compress about the center of the display. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 135: Fft Operation

    In the Math menu, click or tap the Operator drop-down button to select FFT to Figure 9.7 access the menu as shown in Figure 9.7 FFT Operation Menu 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 136: Figure 9.8 Fft Operation Window

    • Span-Center (frequency span to center frequency): Span specifies the frequency range represented by the width from the frequency at the left side of 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 137: Table 9.1 Window Function

    Poorest amplitude resolution Rectangular Sine waveforms with the same Similar to the situation when no amplitudes and rather similar window is applied. frequencies 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 138 Peak Search Click or tap the icon at the right side of Peak Search to enter the peak search menu, as shown in the figure below. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 139: Figure 9.9 Peak Search

    Select the desired location to save the file and then click or tap Save to save the peak search results. Clicking or tapping the icon at the right side of Peak Search can close the peak search menu. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 140: Logic Operation

    "1"; otherwise it is logic "0". The Table 9.2 Logic results of logic XOR operation of two binary bits are shown in Operation 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 141: Figure 9.11 Operation Result Display Window

    Click or tap the drop-down button of SourceA or SourceB to select CH1, CH2, CH3, or CH4. When a source channel is selected, the selected channel automatically switches to the ON state. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 142 Click or tap the threshold input field of the specified channel and use the pop-up numeric keypad to set the threshold or use the corresponding multipurpose knob to set the value. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 143: Digital Filter

    • BandStop only allows the signals whose frequencies are lower than the current lower limit frequency or higher than the current upper limit frequency to pass. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 144: Figure 9.12 Digital Filter Menu

    Click or tap the Source drop-down button to select from CH1~CH4 or Ref1~Ref10. When a source channel is selected, the selected channel automatically switches to the ON state. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 145 Invert is used to enable or disable the inverted display of the waveform. For setting To Invert a Waveform methods, refer to descriptions in Waveform This oscilloscope provides Main and Zoom for the measurement range. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 146 Center: When the vertical scale is changed, the math operation waveform will expand or compress about the center of the display. Grid To Set the Screen Grid For setting methods, refer to the descriptions in 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 147: Measurements

    Measurements Measurements DHO1000 series oscilloscope provides the quick measurements after "Auto" is selected, auto measurements for 41 waveform parameters, as well as the cursor measurement function. 10.1 Auto Scale When the oscilloscope is correctly connected and has detected a valid input signal, click or tap the function navigation icon >...
  • Page 148: Auto Measurements

    (too large or too small), then the measurement results are invalid, and "*****" is displayed on the screen. Please re-input the signal or set the signal. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 149: Time Parameters

    Tvmin: indicates the time that corresponds to the minimum value of the waveform (Vmin). The default values for threshold upper limit, threshold middle value, and threshold low limit are 90%, 50%, and 10%, respectively. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 150: Count Values

    Rising Edge Count It is specified as the number of rising edges that rise from under the threshold lower limit to above the threshold upper limit. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 151 It is specified as the number of falling edges that fall from above the threshold upper limit to below the threshold lower limit. Falling Edge Count = n Threshold Upper Limit Threshold Lower Limit 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 152: Delay And Phase Parameters

    6. Phase(f-f): indicates the phase deviation between the threshold middle values of the falling edge of Source A and that of Source B. The phase formula is as follows: DelayA PhaseA × ° Period sourceA 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 153: Voltage Parameters

    Source A and Source B can be any channel among CH1~CH4 and Math1~Math4. • The default threshold middle value is 50%. 10.2.1.4 Voltage Parameters Overshoot Vmax Vupper Vtop Vmid Vbase Vlower Vmin Preshoot Figure 10.3 Voltage Parameters 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 154 14. Preshoot: indicates the ratio of the difference between the minimum value and the base value of the waveform to the amplitude value. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 155: Other Parameters

    You can also refer to to enable all measurements. • Vertical: Vmax, Vmin, Vpp, Vtop, Vbase, Vamp, Vupper, Vmid, Vlower, Vavg, VRMS, Per. VRMS, Overshoot, Preshoot, Area, Per.Area, and AC.RMS. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 156: Figure 10.4 Vertical Measurement Items

    Horizontal: Period, Frequency, Rise Time, Fall Time, +Width, -Width, +Duty, - Duty, Positive Pulse Count, Negative Pulse Count, Rising Edge Count, Falling Edge Count, Tvmax, Tvmin, +Slew Rate, and -Slew Rate. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 157: Figure 10.5 Horizontal Measurement Items

    Figure 10.5 Horizontal Measurement Items • Other: Delay (r-r), Delay (r-f), Delay (f-r), Delay (f-f), Phase (r-r), Phase (r-f), Phase(f-r), and Phase (f-f). Figure 10.6 Other Measurement Items 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 158: Measurement Settings

    Click or tap the input field of Upper and use the pop-up numeric keypad to set the upper limit of the measurement or use the corresponding multipurpose 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 159 Amplitude Measurement Method Click or tap Auto or Manual as the amplitude measurement method, which affects the measurement method for the top and base values 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 160 You can also use the corresponding multipurpose knob to set the value. Its range is from 2 to 100,000. Its default value is 1000. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 161: Remove The Measurement Results

    X Cursor X cursor is a vertical solid/dotted line that is used to make horizontal adjustments. It can be used to measure time (s) and frequency (Hz). 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 162 AX: indicates the X value at Cursor A. • AY: indicates the Y value at Cursor A. • BX: indicates the X value at Cursor B. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 163: Manual Mode

    Figure 10.9 Manual Mode Setting Menu Select the Measurement Source Click or tap the drop-down button of Source to select the desired channel (None, CH1~ CH1~CH4, or Math1~ Math4). 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 164 Measurement Example Measure the period of a sine wave by using the manual cursor measurement and auto measurement respectively. The measurement results are both 8 μs. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 165: Track Mode

    In the Cursors menu, click or tap Track for the Mode item to enable the Track cursor measurement function. The measurement results are displayed in the "Result" bar at the right section of the screen. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 166: Figure 10.11 Track Mode Setting Menu

    Click or tap the input field of BX and use the pop-up numeric keypad to set the horizontal position of Cursor B (X cursors). Its adjustable range is limited within the screen. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 167: Figure 10.12 Track Measurement (Before Horizontal Expansion)

    Figure 10.13 Figure 10.12 Track Measurement (before Horizontal Expansion) 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 168: Xy Mode

    In the Cursors menu, click or tap XY for the Mode item to enable the XY cursor measurement function. The measurement results are displayed in the "Result" bar at the right section of the screen. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 169: Figure 10.14 Xy Mode Setting Menu

    To configure the multipurpose knobs, please refer to During the adjustment, the measurement results will change accordingly. The adjustable range is limited within the screen. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 170 In the "Result" bar at the right side of the screen, drag the "Cursors" label to the right to simply disable cursor measurements. • Press the front-panel key to disable cursor measurements. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 171: Digital Voltmeter (Dvm) And Frequency Counter

    Click or tap the "DVM" label and then a window is displayed. Click or tap Setting to Figure 11.1 enter the DVM setting menu as shown in . You can click or tap Remove to disable DVM measurements. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 172: Measurement Settings

    Click or tap the Beeper on/off switch to turn on or off the beeper. When the beeper is turned on, you can enable the beeper to sound an alarm when the voltage value is inside or outside the limited range. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 173: Remove The Measurement

    After the counter is enabled, the "Counter" label displaying the counter measurement results appears in the "Result" bar at the right section of the screen, as shown in the figure below. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 174: Measurement Settings

    You can also use the corresponding multipurpose knob to set the value. The range of resolution is from 3 bits to 6 bits. By default, it is 4 bits. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 175: Reset Statistics

    Click or tap Remove to clear the measurement results. The "Counter" label disappears from the "Result" bar accordingly. You can also click or tap Remove in the counter setting menu to disable the function. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 176: Reference Waveform

    To Select the Reference Channel Click or tap the drop-down button of Current to select the reference waveform channel (Ref1-Ref10). By default, Ref1 is enabled. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 177: To Set The Ref Waveform Display

    You can directly click or tap the Up and Down arrow icons at the right side of the input field of VOffset to 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 178: Export And Import Operation

    Input the filename with the pop-up numeric keypad. Parameter For the methods of using the numeric keypad, refer to descriptions in Setting Method • Set the Save Path 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 179 Then click or tap Load to complete the load operation. For details about the disk management operation, Disk Management Store and Load refer to the section in 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 180: Pass/Fail Test

    You can select the source, create mask, and set test result output only after the pass/ fail test function is enabled. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 181: To Select The Source

    To Set the Output Form of the Test Results In Option, you can set what the oscilloscope will execute when test results are detected according to your needs. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 182: To Start Or Stop The Pass/Fail Test Operation

    The "Pass/Fail" result will be displayed in the "Result" bar at the right side of the screen, as shown in the figure below. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 183: To Display The Statistics Of The Test Results

    Click or tap Reset Stat., then the statistics in the "Pass/Fail" label will be reset to • Click or tap Setting, then the PassFail setting menu is displayed. • Click or tap Remove, then the pass/fail function is disabled. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 184: Protocol Decoding

    Figure 14.1 Schematic Diagram of Parallel Decoding In the Decode menu, click or tap the drop-down button of Bus Type to select Parallel, then configure the parameters for Parallel decoding. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 185: Clock Setting (Clk)

    Falling: samples the channel data on the falling edge of the clock. • Both: samples the channel data on the rising edge or the falling edge of the clock. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 186: Bus Setting

    The range of the threshold is related to the current vertical scale and offset. Endian In Endian, select "Invert" or "Normal" as the endian of the bus. Polarity In Polarity, select "Positive" or "Negative" as the data polarity. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 187: Display-Related Settings

    You can also click or tap the icon at the upper-right corner of the table to close the event table. Figure 14.3 Parallel Decoding Event Table 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 188: Rs232 Decoding

    Data Bit: indicates the number of data bits actually contained in each frame of data. • Check Bit: used to check whether the data are properly transmitted. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 189: Source Setting

    When you modify the threshold of the Tx source channel, a dotted line displaying the current threshold level is displayed on the screen. It disappears in about 2 s after you stop modifying the threshold. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 190: To Set Data Package

    Click or tap the drop-down button of Stop Bit to set the stop bits after each frame of data. It can be set to 1 bit, 1.5 bits, or 2 bits. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 191: Display-Related Settings

    When enabled, the event table is displayed as shown in the figure below. You can also click or tap the icon at the upper-right corner of the table to close the event table. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 192: I2C Decoding

    I2C serial bus consists of the clock line (SCL) and the data line (SDA). SCL: samples SDA on the rising or falling edge of the clock. SDA: indicates the data channel. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 193: Source Setting

    Click or tap the input field of SCL Thre, and then use the pop-up numeric keypad to set the threshold of the clock channel. You can also use the corresponding multipurpose knob to set the value. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 194: Display-Related Settings

    When enabled, the event table is displayed as shown in the figure below. You can also click or tap the icon at the upper-right corner of the table to close the event table. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 195: Figure 14.9 I2C Decoding Event Table

    Error expressions in decoding In I2C decoding, the response includes ACK (acknowledgment) and NACK (non- acknowledgment). When NACK is detected after "Write", red error report information ) is displayed. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 196: Spi Decoding

    SPI decoding. Figure 14.11 SPI Decoding Menu Bus Status Click or tap the Bus Status on/off switch to enable or disable the bus decoding. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 197: To Set The Source

    You can perform frame synchronization according to the timeout, and the timeout value must be greater than half of the clock cycle. Click or tap the input 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 198 You can also use the corresponding multipurpose knob to set the value. The setting range is from 4 to 32. By default, it is 8. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 199: Display-Related Settings

    When you adjust the horizontal time base, the waveform displayed on the screen will also change, and the total number of lines containing the decoding information in the event table will also be changed. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 200: Can Decoding

    Copy trig indicates applying the trigger settings to the specified decoding setting. Click or tap Copy Trig to apply the trigger settings to the specified decoding setting. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 201: Signal Configuration

    The sample position is represented by the proportion of "the time from the start of the bit to the sample point" to the "bit time", as shown in the figure below. One Bit Figure 14.14 Sample Position 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 202: Display-Related Settings

    You can also click or tap the icon at the upper-right corner of the table to close the event table. Figure 14.15 CAN Decoding Event Table 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 203: Lin Decoding

    LIN signal protocol version. In the Decode menu, click or tap the drop-down button of Bus Type to select LIN, then configure the parameters for LIN decoding. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 204: Signal Configuration

    You can also self-define the baud rate. • In Parity bit, click or tap "With" or "Without" to select whether the data contain the parity bit. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 205: Display-Related Settings

    You can also click or tap the icon at the upper-right corner of the table to close the event table. Figure 14.17 LIN Decoding Event Table 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 206 Checksum (check sum): expressed in Hex, identified as "Checksum:". When errors occur, it is displayed as a red patch. • Wakeup (wake up symbol): identified as "Wakeup:". 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 207: 15 Multi-Pane Windowing

    Multi-pane Windowing 15 Multi-pane Windowing This series oscilloscope supports multi-pane windowing. You can add multiple windows and result display windows for display and view. Click or tap > Windows to enter the "Add Window" menu. You can also click or tap the Windows button on the toolbar at the top of the screen to enter the menu, as shown in the figure below.
  • Page 208: Waveform Recording And Playing

    Click or tap the Record on/off switch to enable or disable the waveform recording Record function. Before recording the waveform, you can refer to descriptions in Options to set the waveform recording parameters. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 209: Record Options

    2. Frames The recording frames refer to the number of frames that can be recorded actually. After starting the recording operation, the oscilloscope stops the recording 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 210: Play Options

    When ON is selected, the window is minimized, making the display more simplified, easy for observation and operation. The minimized menu is as shown in the figure below. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 211: Figure 16.2 Minimized "Play" Menu

    You can also use the corresponding multipurpose knob to set the value. The available range is from 1 ms to 1 s. 4. Start Frame 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 212 You can also use the corresponding numeric keypad to set the value. The maximum allowed is the number of recorded frames. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 213: Search And Navigation

    Click or tap Search on /off switch to enable or disable the search function. Zoom Mode (Delayed Sweep) Enabling the search function automatically turns on the 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 214 The inverted triangle mark of the selected event turns red like - Click or tap at the upper-right side of the table to open the search menu. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 215: Navigation

    You can assess the Navigation menu in the following ways. • Press the front-panel key to access the menu. • Click or tap Navigate on the toolbar at the upper-right of the interface. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 216: Figure 17.3 Navigation Menu

    After searching events using the Search function, you can select the "Search Event" mode in Navigation menu to navigate to specific events in the marktable. Its setting Figure 17.5 menu is as shown in 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 217: Figure 17.5 "Search Event" Navigation Setting Menu

    Click or tap Start Frame and End Frame input fields to set the start frame and end frame which define the segment range to navigate through. You can also use the corresponding multipurpose knob to set the start/end frame. After you 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 218 The current page/total pages is indicated in Current Page. During segment playing, you are not allowed to change the start/end frame. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 219: Display Control

    The available values are Min, specific values (100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s), and Infinite. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 220: Waveform Intensity

    In the Display setting menu, drag the slider of Grid Brightness to set the grid brightness. The default is 50%, and the range available is from 0% to 100%. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 221: Show Scale

    STOP/RUN button at the upper-right side of the screen or press the front-panel key to stop sampling. If this function is disabled, the last triggered waveform is displayed. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 222: Store And Load

    Choose to select "Save Image" to enter the "Save Image" setting menu. Set the relevant parameters and save the image to the internal or external memory. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 223: Figure 19.1 Image Saving Setting Menu

    Click or tap the input field of File Name to input the file name to be saved with the pop-up virtual keypad. • Set the file path 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 224: Save Wave

    "Save Wave" setting menu. The main settings (e.g. channel on/off state, vertical scale, and horizontal time base) and waveform data of all enabled channel will be saved to the internal or external memory. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 225: Figure 19.2 Waveform Saving Setting Menu

    When a USB device is not connected, the default file path is the "Local Disk"; when a USB device is detected, the path is automatically set to "D:". 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 226: Save Setup

    "*.stp" format. When loading, the stored settings can be recalled. Figure 19.3 Setup Saving Setting Menu Set the file saving parameters • Set the filename 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 227: Binary Data Format (.Bin)

    16 Bytes n Bytes In BIN file format, it contains the following channel data: • CH1 Data • CH2 Data • CH3 Data • CH4 Data 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 228: Table 19.2 File Header

    1. File Header There is only one file header in a binary file. The file header contains the following information. Table 19.2 File Header Two-byte characters, RG, indicating that the file is the RIGOL Cookie binary data file format. Version Two-byte, indicating the file version.
  • Page 229 A 16-byte character array, indicating the date when the file is Date saved. A 16-byte character array, indicating the time when the file is Time saved. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 230: Load A File

    In the storage setting menu, click or tap the Load tab to switch to the load menu. In this menu, you can load the local file to the instrument. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 231: Firmware Upgrade

    This instrument supports local upgrade and online upgrade. Local upgrade 1. In the storage setting menu, click or tap Upgrade to enter the local upgrade setting menu. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 232: Disk Management

    Then click or tap Disk at the lower-left corner of the "Storage" menu to enter the disk management interface, as shown in the figure below. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 233: Figure 19.6 Disk Management Interface

    Click or tap the check box at the right side of the folder, if checked, it is selected, with an icon being displayed. Click or tap the check box again or to deselect it. The check box restores its original state. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 234 Select a specified file or folder, then click or tap Rename to input a new filename or folder name with the pop-up virtual keypad. Then, the rename operation is completed. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 235: System Utility Function Setting

    For each instrument, the MAC address is unique. When assigning the IP address for the instrument, the system uses the MAC address to identify the instrument. VISA Address Displays the VISA address currently used by the instrument. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 236 The range of the first segment (nnn) is from 0 to 223 (except 127), and the range for the other three segments is from 0 to 255. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 237: Basic Settings

    This product supports menus in multiple languages, including the display of the help information, prompt messages, and interface. Click or tap the drop-down button of Language to select the specified system language. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 238 Once the input lock is enabled, the input function is disabled, for which you can no longer use the touch screen, front-panel keys, and knobs to configure the instrument. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 239: About This Oscilloscope

    Indicates the product model. • Serial Number Indicates the serial number, the unique identification for the product. • Firmware Indicates the firmware version number of the product. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 240: Other Settings

    If "ON" is 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 241: Selfcal

    After starting the self-calibration program, click or tap Exit to cancel self- calibration operation at any time. • Click or tap Close to close the self-calibration information window. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 242: Option List

    After setup, click or tap the key at the upper-right corner of the front panel to capture the current screen and save the image with the specified format. The location 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 243 This resets all statistics of the specified function in the Result sidebar at the right side of the screen and starts the measurements again. Waveform Recording Click or tap Record, then Operation is set to "Record". 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 244: Self-Check

    At this time, you can press the front-panel keys to check whether the virtual keys are highlighted. If yes, it indicates that the keys work normally; if no, it indicates that 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 245: Figure 20.4 Touch Screen Test Interface

    Board Test Click or tap Board Test, then the board test interface is displayed. Check whether the status of each module is in good condition. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 246: Remote Control

    SCPI (Standard Commands for Programmable Instruments) commands. This chapter will illustrate how to use the RIGOL Ultra Sigma software to remotely control the instrument via various interfaces. Note: When communicating with the PC via GPIB, the instrument does not support large data transmission operation such as screenshot and waveform reading.
  • Page 247: Remote Control Via Usb

    PC via the USB interface. You can also click USB-TMC to search for the resource. 3. View the device resource The resources found will appear under the "RIGOL Online Resource" directory, and the model number and USB interface information of the instrument will also be displayed.
  • Page 248: Remote Control Via Gpib

    4. View the device resource The resources found will appear under the "RIGOL Online Resource" directory. 5. Control the instrument remotely Right-click the device resource name and select "SCPI Panel Control" to open the remotely command control panel.
  • Page 249: Figure 21.1 Search For The Available Device

    5. View the device resource Click OK to go back to the main interface of Ultra Sigma. The searched instrument resource will be displayed under the directory of "RIGOL Online Resource". 6. Control the instrument remotely Right-click the device resource name. In the displayed menu, select "SCPI Panel Control"...
  • Page 250: Troubleshooting

    (e.g. a magnet), please move the instrument away from the magnet field. d. If the problem still persists, please contact. 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 251: Appendix

    23.2 Appendix B: Warranty RIGOL TECHNOLOGIES CO., LTD. (hereinafter referred to as RIGOL warrants that the product mainframe and product accessories will be free from defects in materials and workmanship within the warranty period. If a product proves defective within the warranty period, RIGOL guarantees free replacement or repair for the defective product.
  • Page 252: Appendix C: Factory Settings

    There is no implied warranty of merchantability or fitness for a particular purpose. Under no circumstances shall RIGOL be liable for any consequential, indirect, ensuing, or special damages for any breach of warranty in any case.
  • Page 253 Show Scale Color Grade Waveform Freeze Measure Threshold Indicator Statistics Count 1,000 Type Source Upper Lower Amplitude Measurement Auto Method Region Main Cursor Mode Manual Source Select 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 254 Beeper Upper Limit Lower Limit Save Image Format *.png Invert Color Color Header Overlay Save Wave Data Source Screen Format *.bin Save Setup File Type *.stp 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 255 Format (to save file) *.pf File Name RigolDS Aux Output Pulse 1 μs Output Event Fail Polarity Positive Error Action Power Analysis Display 1000 Power Quality 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 256 100 ms Math Operation Invert Expand Display Label Grid FULL Operation SourceA SourceB Scale 500 mV Offset Operation SourceA SourceB Scale 500 mV Offset A×B Operation 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 257 1.8 dBV Table Order Amp Order A&&B Operation SourceA SourceB Wave Size Medium Offset 0 div Sensitivity 300 mdiv Thre.CH1 Thre.CH2 Thre.CH3 Thre.CH4 A||B Operation SourceA SourceB 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 258 300 mdiv Thre.CH1 Thre.CH2 Thre.CH3 Thre.CH4 Intg Operation Source Scale 500 mV*s Offset 0 V*s Bias Diff Operation Source Scale 500 mV/s Offset 0 V/s Smooth Sqrt 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 259 Scale 500 mV Offset ωc 4 kHz High Pass Operation Source Scale 500 mV Offset ωc 4 kHz Band Pass Operation Source Scale 500 mV Offset 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 260 50 mV VOffset Label REF1 Label Display Color Orange Decode Bus Type Parallel Bus Status Format Label Event Table Parallel Threshold Endian Invert Polarity Positive RS232 Threshold 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...
  • Page 261 1 μs Polarity Positive Width Endian Source Threshold Baud Rate 19.2 kbps Parity Without Version Both Source Threshold Signal CAN_L Baud 1 Mbps Sample Position FlexRay 1.888.610.7664 www.calcert.com sales@calcert.com Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 User Guide...