Power Analysis (Optional)
12.2
Ripple
Power ripple is an important specification to consider for evaluating DC power supply.
It indicates the ripple quantity of the output DC voltage. The ripple analysis can
measure the current value, average value, minimum value, maximum value, standard
deviation, and count value of the ripples on the power output terminal.
NOTE
A 1X probe (e.g. PVP2150 and PVP2350) is recommended for the ripple measurements.
•
Select the Analysis Type
Click or tap the Type drop-down button to select "Ripple".
•
Set the Ripple Analysis Source
Click or tap the Source drop-down button to select the source (CH1~CH4).
•
Set Count
Click or tap the Cnt input field to set the number of statistical times for ripple
analysis with the numeric keypad. You can also use the corresponding
multipurpose knob to set the value. Its range is from 2 to 5000. By default, it is
1000.
•
Reset
Click or tap Reset Stat. to clear the current data and execute statistics on the
measurement results again.
•
Open or Close the Results Display Window