  • Page 2 • RIGOL products are covered by P.R.C. and foreign patents, issued and pending. • RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at the company's sole decision. • Information in this publication replaces all previously released materials.
  • Page 3: Table Of Contents

    Test Procedures......................5.3.4 Test Record From....................... DC Gain Accuracy Test......................16 5.4.1 Specification........................ 5.4.2 Test Connection Diagram..................5.4.3 Test Procedures......................5.4.4 Test Record From....................... DC Offset Accuracy Test....................5.5.1 Specification........................ Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 4 Test Connection Diagram..................5.6.3 Test Procedures......................5.6.4 Test Record From....................... Time Base Accuracy Test....................25 5.7.1 Specification........................ 5.7.2 Test Connection Diagram..................5.7.3 Test Procedures......................5.7.4 Test Record From....................... Appendix: Test Record Form....................28 Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 5 Figure 5.3 DC Gain Accuracy Test Connection Diagram ..........Figure 5.4 DC Offset Accuracy Test Connection Diagram ......... Figure 5.5 Bandwidth Test Connection Diagram ............Figure 5.6 Time Base Accuracy Test Connection Diagram ........Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 6 List of Tables Table 2.1 Specifications ......................Table 5.1 Test Devices ......................11 Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 7: Safety Requirement

    Do Not Insert Objects Into the Air Outlet. Do not insert objects into the air outlet, as doing so may cause damage to the instrument. • Use Proper Fuse. Please use the specified fuses. Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 8: Safety Notices And Symbols

    Do Not Operate With Suspected Failures. If you suspect that any damage may occur to the instrument, have it inspected by RIGOL authorized personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.
  • Page 9 It calls attention to an operation, if not correctly performed, could result in damage to the product or other devices connected to the product. Safety Symbols on the Product: Hazardous Safety Warning Protective Earth Chassis Ground Test Ground Voltage Terminal Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 10: Specifications

    Specifications This chapter lists the technical specifications of the DHO1000 series oscilloscope. DHO1000 series digital oscilloscope achieves 2 GSa/s sample rate, a fast waveform capture rate of 1,500,000 wfms/s with the UltraAcquire mode, 100 Mpts memory depth, 12-bit vertical resolution, all combined with excellent noise floor performance and vertical accuracy to meet your requirements for more accurate measurements, bringing extraordinary T&M experience for you.
  • Page 11 Specifications Overview of the DHO1000 Series Technical Specifications Bandwidth 20 MHz, FULL; selectable for each channel Limit 2-channel model: 2 GSa/s (single channel ), 1 GSa/s (all Max. Analog channels Channel 4-channel model: 2 GSa/s (single channel ), 1 GSa/s (half...
  • Page 12: Document Overview

    RIGOL DHO1000 series digital oscilloscope. For the operation methods mentioned in the test procedures, refer to User Guide of this product. http:// For the latest version of this manual, download it from the official website of RIGOL ( www.rigol.com Publication Number...
  • Page 13 Document Overview Model Max. Analog Bandwidth No. of Analog Channels DHO1072 70 MHz DHO1074 70 MHz DHO1102 100 MHz DHO1104 100 MHz DHO1202 200 MHz DHO1204 200 MHz Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 14: Overview

    Make sure that the oscilloscope has been warmed up or operating for more than 30 minutes before the self-calibration. 1. Disconnect all the input channels. 2. In "Utility" menu, click or tap SelfCal. The self-calibration interface will be displayed as shown below. Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 15: Test Result Record

    It is recommended that users photocopy the test record form before each test and record the test results in the copy so that the form can be used repeatedly. Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 16: Specifications

    Overview Specifications The specification of each test item is provided in this manual. For other technical DHO1000 Data Sheet (download it from RIGOL website: parameters, refer to www.rigol.com All the specifications are valid only when the oscilloscope has been warmed up for more than 30 minutes.
  • Page 17: Performance Verification Test

    Performance Verification Test This chapter takes DHO1104 as an example to illustrate the performance verification test methods and procedures of DHO1000 series digital oscilloscope. This manual recommends the following test devices. You can also use other devices that fulfill the Table 5.1 Test Devices...
  • Page 18: Test Procedures

    64 μV 68 μV 77 μV 1 mV/div Full BW 68 μV 76 μV 77 μV 2 mV/div Full BW 74 μV 82 μV 95 μV 5 mV/div Full BW Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 19: Impedance Test

    Before connecting, disconnecting, or moving the test hookup, disable the output of the signal generator to avoid causing the dangerous voltage. 1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in Figure 5.1 Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 20: Test Record From

    0.99 MΩ to 1.01 MΩ 50 mV/div 0.99 MΩ to 1.01 MΩ 500 mV/div 0.99 MΩ to 1.01 MΩ Input Capacitance Test 5.3.1 Specification Input Capacitance Specification 19 pF ± 3 pF Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 21: Test Connection Diagram

    4. Adjust the vertical scale of CH1 to 500 mV/div; read and record the capacitance measured. 5. Turn off CH1. Measure the capacitances of CH2, CH3, and CH4 respectively using the method above and record the measurement results. Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 22: Test Record From

    1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in Figure 5.3 2. Set the impedance of Fluke 9500B to 1 MΩ. 3. Output a DC signal with +3 mVdc voltage (Vout1) via Fluke 9500B. Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 23 Adjust the output voltage of Fluke 9500B to 3 x the current vertical scale and -3 x the current vertical scale respectively. c. Repeat Step 3-7 and record the test results. Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 24: Test Record From

    2 V/div ≤2% 5 V/div ≤2% 10 V/div ≤2% 1 mV/div ≤2% 2 mV/div ≤2% 5 mV/div ≤2% 10 mV/div ≤2% 20 mV/div ≤2% 50 mV/div ≤2% 100 mV/div ≤2% Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 25: Dc Offset Accuracy Test

    DC Offset Accuracy ≤200 mV/div (±0.1 div ± 2 mV ± 1.5% of the offset value) Specification >200 mV/div (±0.1 div ± 2 mV ± 1.0% of the offset value) Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 26: Test Connection Diagram

    Click or tap the trigger label at the top of the screen. Set the trigger source to AC Line. g. Adjust the trigger level to prevent the signals from being triggered by mistake. 4. Set the output of Fluke 9500B to -1 V. Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 27: Test Record From

    -1.01 V -982.95 mV 500 μV/div -2.05 mV 2.05 mV CH3, 20 MHz BW 500 μV/div -1 V -982.95 mV 1.01 V 200 mV/div 10 V -10.17 V -9.92 V Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 28: Bandwidth Test

    [1]: Amplitude Loss (dB) = 20 x lg (Vrms2/Vrms1); wherein, Vrms1 is the measurement result of amplitude effective value at 1MHz and Vrms2 is the measurement result of amplitude effective value at full bandwidth. Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 29: Test Connection Diagram

    In this menu, you can set the horizontal time base to 500 ns/ div. f. Set the horizontal position to 0 s. Close the Horizontal menu. g. Set the trigger level to 0 V. Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 30: Test Record From

    -3 dB to 3 50 mV/div -3 dB to 3 50 mV/div -3 dB to 3 500 mV/div -3 dB to 3 50 mV/div -3 dB to 3 500 mV/div NOTE Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 31: Time Base Accuracy Test

    1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in Figure 5.6 2. Output a Sine signal with 10 MHz frequency and 1 Vpp amplitude via Fluke 9500B. 3. Configure the oscilloscope: Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 32 1 ns/1 ms = 1 ppm. 6. Calculate the time base accuracy limit by using the following formula: ±(5 ppm + 1 ppm/year x Number of Years that the Instrument Has Been Used). Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 33: Test Record From

    [2]: For the number of years that the instrument has been used, please calculate it according to the date in the verification certificate provided when the instrument leaves factory. Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 34: Appendix: Test Record Form

    Appendix: Test Record Form Appendix: Test Record Form RIGOL DHO1000 Series Digital Oscilloscope Performance Verification Test Record Form Model: ________________ Tested by: ________________ Test Date: ________________ Random Noise Test Record Form Maximum Vertical Test Pass/ Bandwidth DHO1072/ DHO1102/ DHO1204/ Scale...
  • Page 35 20 mV/div ≤2% 50 mV/div ≤2% 100 mV/div ≤2% 200 mV/div ≤2% 500 mV/div ≤2% 1 V/div ≤2% 2 V/div ≤2% 5 V/div ≤2% 10 V/div ≤2% 1 mV/div ≤2% Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 36 10 V -10.17 V -9.92 V MHz BW 200 mV/div -22.00 mV 22.00 mV 200 mV/div -10 V 9.92 V 10.17 V 1 V/div 20 V -20.30 V -19.69 V Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 37 1 V/div -20 V 19.69 V 20.30 V 5 V/div 100 V -101.50 V -98.49 V 5 V/div -502.00 mV 502.00 mV 5 V/div -100 V 98.49 V 101.50 V Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...
  • Page 38 [2]: For the number of years that the instrument has been used, please calculate it according to the date in the verification certificate provided when the instrument leaves factory. Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. DHO1000 Performance Verification...