Duration Trigger; Figure 8.11 Duration Trigger - Rigol HDO4000 Series User Manual

Digital oscilloscope
Hide thumbs Also See for HDO4000 Series:
Table of Contents

Advertisement

Trigger Parameter Setting
You can refer to
noise rejection under this trigger type.
Trigger Level
Click or tap the Level input field to set the trigger level with the pop-up numeric
keypad. You can also use the front-panel trigger level knob or the corresponding
multipurpose knob to set the trigger level. For details, refer to descriptions in
Level
. The current trigger level is displayed in the trigger information label at the top
of the screen.
8.7.6

Duration Trigger

In duration trigger, the oscilloscope identifies a trigger condition by searching for the
duration of a specified pattern. This pattern is a logical "AND" combination of the
channels. Each channel can be set to 1 (high), 0 (low), or X (don't care). The
instrument triggers when the duration (∆T) of this pattern meets the preset time, as
shown in the figure below.
Trigger Type
Click or tap the Type drop-down button to select "Duration".
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Trigger Holdoff
Noise Rejection
and
∆T
CH1
CH2
CH3
CH4

Figure 8.11 Duration Trigger

Triggering the Oscilloscope
to set the trigger holdoff and
∆T
Pattern:HLHL
HDO4000 User Guide
Trigger
85

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hdo4204Hdo4404Hdo4804

Table of Contents