System Combinations - Zeiss S7 / OPMI VISU 160 Instructions For Use Manual

Table of Contents

Advertisement

S7 / OPMI VISU 160

System combinations

CAUTION
Version 6.0
G-30-1528-en
The device S7 / OPMI VISU 160 can be extended with the accessories offered
by ZEISS (see ordering data/accessories) to form an already tested system. This
system combination has been approved by ZEISS.
In accordance with EN 60601-1:2006, section 16.2, the power outlet socket
of this system is a multiple connector that is intended for an ME system. Con-
necting electrical devices not approved by ZEISS can lead to a reduced level
of safety on the ME system. When configuring your ME system (use of the
power outlet socket), make sure to observe the requirements of IEC 60601-
1:2005 section 16, and note the following:
• Never place multiple connectors on the floor.
• Never connect any additional multiple connectors.
• Only connect compatible components to the system.
• Ensure that you do not exceed the max. permissible load capacity of the
multiple connectors.
• Use multiple connectors only for components which are part of the sys-
tem.
Any additional equipment, or equipment other than specified above, which is
connected to this system must demonstrably comply with the applicable
standards or directives (e.g. IEC 60950 Standards for data processing equip-
ment). Furthermore, all configurations must comply with the normative re-
quirements for medical systems (see IEC 601-1-1:1990 or IEC 60601-1:2005,
Section 16).
Anyone connecting additional equipment to the approved device combina-
tion is a system configurer and as such responsible for compliance of the
system with the normative requirements for systems. Please note that local
legislation takes priority over the above-mentioned normative requirements.
If you have any questions, please contact your local dealer or ZEISS Service.
System Data