Controls And Connections - Zeiss S7 / OPMI VISU 160 Instructions For Use Manual

Table of Contents

Advertisement

System Overview

Controls and connections

-
+
Balance
Page 74
Suspension arm
1 Lamp housing of the halogen light source
The lamp module contains a backup lamp which is automatically swung
into the illumination beam path when the first lamp fails.
2 Friction
Knob for adjusting the friction of the surgical microscope's rotation.
3 Coupling
for easy mounting of the surgical microscope.
4 Friction
Knob for adjusting the friction of the suspension arm's upward/downward
movement.
5 Clamping screw for limiting downward movement
You can limit the downward movement of the suspension arm by adjust-
ing a stop.
6 Balance setting
Knob for setting the spring force for balancing. After mounting the sur-
gical microscope including all accessories, adjust the balance setting of the
suspension arm using this screw. The balance setting procedure is de-
scribed in the chapters "Preparations for use"/"Adjusting the system".
S7 / OPMI VISU 160
Version 6.0
G-30-1528-en