S7 / OPMI VISU 160
Causes of malfunctions and remedies
Problem
No function at all.
Surgical field illumination too
bright.
Insufficient surgical field illu-
mination.
Version 6.0
G-30-1528-en
For your safety
• This system is a high-grade technological product. To ensure optimum
performance and safe working order, we recommend having it checked
by our service representative as part of regular scheduled maintenance.
If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions", attach a sign to the system
stating it is out of order and contact our service representative.
Possible cause
Power plug of the stand not
inserted.
Power switch of suspension
system not switched on.
Automatic circuit breaker in
power switch of suspension
system has been activated.
Line power failure.
Brightness level set too high.
Brightness level set too low.
Halogen lamp not properly
plugged into lamp mount.
Defective light guide (illumi-
nation not uniform).
What to do in the event of malfunctions
Remedy
Plug in the power cord.
Press power switch.
Green indicator light in the
power switch must be lit.
Press power switch again.
Contact in-house electrician.
Adjust brightness on suspen-
sion system or foot control
panel.
Switch off illumination on
suspension system.
Illuminate surgical field using
an OR illuminator.
Contact Service.
Adjust brightness on suspen-
sion system or foot control
panel.
Properly push halogen lamp
into lamp mount.
Contact Service.
Light guide probably needs to
be changed.
See
Page 78
Page 78
Page 78
Page 168
Page 161