Preparations for use
Adjusting the eyecups
CAUTION
Page 138
 Now, the microscope is adjusted in such a way that the image is sharply
defined over the complete magnification range without any renewed fo-
cusing during the magnification setting.
If this is not the case, repeat the procedure.
• Adjust the eyecups of the eyepieces in such a way that the entire visual
field can be seen.
Viewing with eyeglasses: Rotate eyecups in.
Viewing without eyeglasses: Rotate eyecups out.
Adjusting the working distance and magnification
• Position the microscope above the surgical field so as to ensure conven-
ient work.
• Roughly focus the object by changing the distance between microscope
and object.
• Set the microscope to maximum magnification and adjust the fine focus-
ing system until the image is sharply defined.
• Set the microscope to the required magnification. The focal plane is re-
tained in this process.
Risk of injury to the patient caused by parts falling down!
Due to collisions or incorrect installation, parts may fall down and injure the
patient.
• Check the correct and secure installation of all parts before using the de-
vice.
• Check that the floor stand is secured in a steady position and can be piv-
oted without causing collisions.