What to do in the event of malfunctions

Lamp failure of the halogen light source

NOTE
Page 160
Overheating of the lamp module!
If ventilation grids are covered e.g. by drapes, this may lead to overheating
of the lamp housing and deactivation of the lamp.
• Never cover the ventilation grids.
• Lamps switched off due to overheating will be automatically switched on
again when they have cooled down.
The lamp housing contains a backup lamp which automatically swings into
the illumination beam path when the first lamp fails. Open flap (2) indicates
that the backup lamp is operative. If the automatic swing-in function is defec-
tive, you can also swing in the backup lamps manually.
• Press button (3) to manually swing the backup lamp into the beam path.
• Check that ventilation grid (1) is not covered.
• Exchange the defective lamp at the first opportunity (see Page 168).
• If the backup lamp is defective and replacement is not possible, continue
surgery using an external OR illuminator.
S7 / OPMI VISU 160
1
2
3
Version 6.0
G-30-1528-en