S7 / OPMI VISU 160

Sterilization

CAUTION

Disinfection

NOTE
Version 6.0
G-30-1528-en
Risk of infection caused by contaminated asepsis accessories and
drapes!
• Only use sterile asepsis accessories and drapes suitable for the system!
• The products contained in the asepsis sets must be cleaned, disinfected
and sterilized before each use. This also applies to the first use after de-
livery.
• Make sure that the operating staff is briefed about the sterile operation
of the system.
The asepsis sets available from ZEISS contain rubber caps and hand grips that
can be sterilized in autoclaves. For detailed information on sterilization, please
see the enclosed instructions "Preparation of resterilizable products" for the
respective asepsis set.
You can also use disposable sterile drapes to cover the system.
• When fitting the sterile drape, ensure there is sufficient slack to allow for
movement of the microscope carrier and surgical microscope.
• Do not cover the ventilation opens - this ensures the lamps are cooled ad-
equately and no lamp fails.
Risk of damaging the paint on the device!
• Use a disinfectant based on aldehyde or alcohol. The addition of quater-
nary compounds is acceptable.
To avoid damaging the surfaces, do not use any disinfecting components
other than those specified below.
The maximum concentrations are:
For alcohol (tested with 2 propanol): 60%
For aldehyde (tested with glutaraldehyde): 2%
For quaternary compounds (tested with DDAC): 0.2%
Care and Maintenance
Page 165