Preparations for use
Connecting the suspension system - S7 ceiling mount with
lifting column
Page 122
• Check the voltage indicated at (2).
The suspension system is set at the factory to the rated voltage used in the
country of destination. The rated voltage indicated at window (2) must
correspond to the rated voltage available on the site of installation.
• If the rated voltage has not been correctly set, adjust the sliding switch (2)
using a suitable tool.
• Plug the connector of the foot control panel (option) into connector (1) of
the suspension system or into the wall socket (option). Item (3) shows the
on-wall version, item (4) the in-wall version of the wall socket.
• Before using a cordless foot control panel (FCP WL), ensure that the bat-
teries of the FCP WL are fully charged. The system may malfunction if the
FCP WL is not provided an adequate level of power.